The objectives of the present study were (1) to compare the stability of delayed loaded (DL) and immediately loaded (IL) ITI SLA implants during the first 3 months of the healing period using resonance frequency analysis (RFA) and (2) to determine the factors that affect implant stability during the healing period.
To compare implant stability, RFA was performed on two groups of patients (12 patients received 25 IL implants and 47 patients received 79 DL implants) with a total 104 ITI SLA implants. Implant stability was measured directly by RFA at implant placement and consecutively once a week for 12 weeks. Statistical analyses were carried out to study implant stability differences between IL and DL groups.
One of the 25 implants in the IL group failed, and no implant was lost in the DL group. Implant stability between the IL and DL groups showed a statistically significant difference (P<0.05). The mean implant stability quotient of all measured implants from implant insertion to 12 weeks was 72.88 +/- 5.39 for the DL and 75.86 +/- 3.60 for the IL types. The lowest stability was at 4 weeks for DL implants (mean: 71.58 +/- 5.11) and 2 weeks for IL implants (mean: 71.33 +/- 2.97). In both groups, bone types I and II showed higher implant stability than bone type III (P<0.05).
The findings of this study indicate that differences in osseointegration between IL and DL implants may be predicted according to differential implant stability.
本研究的目的是(1)比较延迟加载(DL)和即刻加载(IL) ITI SLA 种植体在愈合期的前 3 个月使用共振频率分析(RFA)的稳定性,(2)确定愈合期影响种植体稳定性的因素。
为了比较种植体稳定性,对两组患者(12 名患者接受 25 个 IL 种植体,47 名患者接受 79 个 DL 种植体)共 104 个 ITI SLA 种植体进行了 RFA。种植体稳定性通过 RFA直接测量,在种植体植入时和随后的 12 周内每周测量一次。对 IL 和 DL 组之间的种植体稳定性差异进行了统计分析。
IL 组的 25 个种植体中有 1 个失败,DL 组无种植体丢失。IL 和 DL 组之间的种植体稳定性显示出统计学上的显著差异(P<0.05)。从种植体插入到 12 周时,所有测量种植体的平均种植体稳定性指数为 DL 组 72.88 +/- 5.39,IL 组 75.86 +/- 3.60。DL 种植体的最低稳定性在第 4 周(平均:71.58 +/- 5.11),IL 种植体在第 2 周(平均:71.33 +/- 2.97)。在两组中,骨类型 I 和 II 显示出比骨类型 III 更高的种植体稳定性(P<0.05)。
本研究的结果表明,根据种植体稳定性的差异,可以预测 IL 和 DL 种植体之间的骨整合差异。