Sun Yaxun, Zhang Ping, Li Xuebin, Guo Jihong
Department of Cardiac Electrophysiology, People's Hospital of Peking University, Xizhimen South Street, Beijing 100044, China.
Pacing Clin Electrophysiol. 2010 Jan;33(1):113-6. doi: 10.1111/j.1540-8159.2009.02538.x. Epub 2009 Sep 30.
A 45-year-old man with diagnosis of short QT syndrome underwent successful implantation of dual-chamber implantable cardioverter-defibrillator (ICD) to prevent sudden cardiac death. Ten days after implantation, the patient was treated by inappropriate ICD discharges for frequent T-wave oversensing. After careful reprogramming of the ICD, the T-wave oversensing was eliminated and no T-wave oversensing or inappropriate discharge was documented during 6-month follow-up period.