Suppr超能文献

使用锥形束 CT(CBCT)进行骨盆自适应放疗的剂量计算准确性。

Dose calculation accuracy using cone-beam CT (CBCT) for pelvic adaptive radiotherapy.

机构信息

Department of Radiation Oncology, Good Samaritan Health System, Kearney, NE 68848, USA.

出版信息

Phys Med Biol. 2009 Oct 21;54(20):6239-50. doi: 10.1088/0031-9155/54/20/013. Epub 2009 Oct 1.

Abstract

This study is to evaluate the dose calculation accuracy using Varian's cone-beam CT (CBCT) for pelvic adaptive radiotherapy. We first calibrated the Hounsfield Unit (HU) to electron density (ED) for CBCT using a mini CT QC phantom embedded into an IMRT QA phantom. We then used a Catphan 500 with an annulus around it to check the calibration. The combined CT QC and IMRT phantom provided correct HU calibration, but not Catphan with an annulus. For the latter, not only was the Teflon an incorrect substitute for bone, but the inserts were also too small to provide correct HUs for air and bone. For the former, three different scan ranges (6 cm, 12 cm and 20.8 cm) were used to investigate the HU dependence on the amount of scatter. To evaluate the dose calculation accuracy, CBCT and plan-CT for a pelvic phantom were acquired and registered. The single field plan, 3D conformal and IMRT plans were created on both CT sets. Without inhomogeneity correction, the two CT generated nearly the same plan. With inhomogeneity correction, the dosimetric difference between the two CT was mainly from the HU calibration difference. The dosimetric difference for 6 MV was found to be the largest for the single lateral field plan (maximum 6.7%), less for the 3D conformal plan (maximum 3.3%) and the least for the IMRT plan (maximum 2.5%). Differences for 18 MV were generally 1-2% less. For a single lateral field, calibration with 20.8 cm achieved the minimum dosimetric difference. For 3D and IMRT plans, calibration with a 12 cm range resulted in better accuracy. Because Catphan is the standard QA phantom for the on-board imager (OBI) device, we specifically recommend not using it for the HU calibration of CBCT.

摘要

本研究旨在评估瓦里安锥形束 CT(CBCT)在骨盆自适应放疗中的剂量计算准确性。我们首先使用嵌入 IMRT QA 体模中的 mini CT QC 体模对 CBCT 的 Hounsfield 单位(HU)进行电子密度(ED)校准。然后,我们使用带有环形物的 Catphan 500 进行校准检查。组合 CT QC 和 IMRT 体模提供了正确的 HU 校准,但带有环形物的 Catphan 则不行。对于后者,不仅聚四氟乙烯(Teflon)不是骨的正确替代品,而且插入物也太小,无法为空气和骨提供正确的 HU。对于前者,我们使用了三种不同的扫描范围(6cm、12cm 和 20.8cm)来研究 HU 对散射量的依赖性。为了评估剂量计算准确性,我们获取并注册了骨盆体模的 CBCT 和计划 CT。我们在两套 CT 上创建了单野计划、3D 适形和调强计划。不进行不均匀性校正时,两套 CT 生成的计划几乎相同。进行不均匀性校正后,两套 CT 之间的剂量差异主要来自 HU 校准差异。对于 6MV,单外侧野计划的剂量差异最大(最大 6.7%),3D 适形计划次之(最大 3.3%),调强计划最小(最大 2.5%)。18MV 的差异通常小 1-2%。对于单外侧野,20.8cm 扫描范围的校准可实现最小的剂量差异。对于 3D 和调强计划,12cm 扫描范围的校准可实现更高的准确性。由于 Catphan 是机载成像(OBI)设备的标准 QA 体模,我们特别建议不要将其用于 CBCT 的 HU 校准。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验