Suppr超能文献

充盐水的袖带有助于防止激光引发的聚氯乙烯气管内导管起火。

Saline-filled cuffs help prevent laser-induced polyvinylchloride endotracheal tube fires.

作者信息

Sosis M B, Dillon F X

机构信息

Department of Anesthesiology, Rush-Presbyterian-St. Luke's Medical Center, Chicago, Illinois 60612.

出版信息

Anesth Analg. 1991 Feb;72(2):187-9. doi: 10.1213/00000539-199102000-00008.

Abstract

To determine whether the filling of tracheal tube cuffs with saline would decrease their combustibility during laser surgery, 20 polyvinylchloride tracheal tubes were studied. The cuffed end of each tracheal tube was inserted into the neck of an empty flask, and the tube and flask were flushed with oxygen for 5 min before cuff inflation. Ten tracheal tubes had their cuffs inflated with air, and 10 were inflated with saline. A Lasersonics LS880 CO2 laser, set to 5 W for five of each of the two types of filled cuffs and to 40 W for the other pair of five tubes, was fired continuously at the cuffs for up to 1 min. No combustion occurred at the 5-W setting. The times to cuff perforation when the laser was set at 5 W were (mean +/- SD) 1.00 +/- 0.83 and 4.21 +/- 3.91 s for the air- and saline-filled cuffs, respectively, a difference that was not statistically significant. The time to deflation of the saline-filled cuff (104.6 +/- 67.5 s) was, however, significantly longer than that of the air-filled cuff (2.59 +/- 1.97 s). When the tracheal tube cuffs were exposed to 40-W laser radiation, the cuff and adjacent tube shaft ignited in all cases when the cuffs were inflated with air, but only in one of five cases when the cuffs were filled with saline (P less than 0.05). The filling of tracheal tube cuffs with saline provides simple, moderately effective partial protection of the cuff of endotracheal tubes during CO2 laser airway surgery.

摘要

为了确定气管导管套囊充入生理盐水是否会降低其在激光手术中的可燃性,对20根聚氯乙烯气管导管进行了研究。将每根气管导管的带套囊端插入空烧瓶颈部,在套囊充气前,用氧气冲洗导管和烧瓶5分钟。10根气管导管的套囊充入空气,10根充入生理盐水。一台Lasersonics LS880二氧化碳激光仪,对两种充入不同物质的套囊各五根设置为5W,对另外五根设置为40W,持续照射套囊长达1分钟。在5W设置下未发生燃烧。当激光设置为5W时,充空气和充生理盐水的套囊穿孔时间分别为(平均值±标准差)1.00±0.83秒和4.21±3.91秒,差异无统计学意义。然而,充生理盐水的套囊放气时间(104.6±67.5秒)明显长于充空气的套囊(2.59±1.97秒)。当气管导管套囊暴露于40W激光辐射时,套囊充空气时所有情况下套囊及相邻的导管轴均会点燃,但套囊充生理盐水时,仅五例中有一例点燃(P<0.05)。在二氧化碳激光气道手术中,气管导管套囊充入生理盐水可为气管内导管套囊提供简单、中度有效的部分保护。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验