血液对耐激光气管导管燃烧的影响。

The effect of blood on laser-resistant endotracheal tube combustion.

作者信息

Sosis M B, Pritikin J B, Caldarelli D D

机构信息

Department of Anesthesiology, Rush-Presbyterian-St. Luke's Medical Center, Chicago, IL 60612.

出版信息

Laryngoscope. 1994 Jul;104(7):829-31. doi: 10.1288/00005537-199407000-00008.

Abstract

The protection afforded against CO2 laser-induced combustion by five different types of tracheal tubes or protective foil wraps was evaluated. They were compared before and after the application of human blood to their external surfaces. The tracheal tubes tested were polyvinylchloride (PVC) tubes wrapped with Venture copper (Cu) foil tape, 3M aluminum (Al) foil tape, and the Laser-Guard protective coating. The Xomed Laser Shield II and Mallinckrodt Laser-Flex tracheal tubes were also tested. A CO2 laser set to 38 W in the continuous mode was directed at the shaft of the tracheal tube under study, which had 5 L/min of oxygen flowing through it. The laser was actuated for 90 seconds or until combustion or melting occurred. The copper foil-wrapped and aluminum foil-wrapped PVC tracheal tubes were unaffected by 90 seconds of laser fire in five trials with each type of tape. However, the application of blood to the foil wrapped PVC tracheal tube shafts resulted in combustion in 3 of 5 copper foil-wrapped tubes and melting of the underlying tracheal tube shaft in 3 of 5 aluminum foil-wrapped PVC tracheal tubes. Blood did not affect the protection afforded by the Laser-Guard coating when it was applied to the shafts of PVC tracheal tubes. Similarly, the Xomed Laser Shield II tracheal tube's shaft offered good protection from the laser both before and after application of blood. Combustion occurred in 1 of 4 Mallinckrodt Laser Flex tracheal tubes studied prior to the application of blood. The application of blood resulted in almost immediate combustion in all 4 Mallinckrodt Laser Flex tracheal tubes tested. The presence of blood on the surface of metallic foil-wrapped or special tracheal tubes may make laser-induced combustion more likely during airway surgery. However, the Laser-Guard protective coating and the Xomed Laser-Shield II tracheal tube provide good protection even when covered with blood.

摘要

评估了五种不同类型的气管导管或防护箔包裹物对二氧化碳激光诱导燃烧的防护效果。在其外表面涂抹人血前后对它们进行了比较。测试的气管导管有包裹了Venture铜(Cu)箔带、3M铝(Al)箔带的聚氯乙烯(PVC)管以及Laser-Guard防护涂层。还测试了Xomed Laser Shield II和Mallinckrodt Laser-Flex气管导管。一台设置为连续模式、功率38W的二氧化碳激光对准正在研究的气管导管的管身,管内有5L/min的氧气流过。激光启动90秒或直至发生燃烧或熔化。每种类型的胶带在五次试验中,包裹铜箔和铝箔的PVC气管导管在90秒的激光照射下均未受影响。然而,在包裹箔的PVC气管导管管身涂抹血液后,5根包裹铜箔的导管中有3根发生燃烧,5根包裹铝箔的PVC气管导管中有3根其下方的气管导管管身熔化。血液涂抹在PVC气管导管管身时,并未影响Laser-Guard涂层提供的防护。同样,Xomed Laser Shield II气管导管的管身在涂抹血液前后均能很好地抵御激光。在涂抹血液之前,所研究的4根Mallinckrodt Laser Flex气管导管中有1根发生燃烧。涂抹血液后,所有4根测试的Mallinckrodt Laser Flex气管导管几乎立即发生燃烧。在金属箔包裹的或特殊的气管导管表面存在血液,可能会使气道手术期间激光诱导燃烧的可能性增加。然而,Laser-Guard防护涂层和Xomed Laser-Shield II气管导管即使被血液覆盖也能提供良好的防护。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验