Suppr超能文献

三种不同电子根尖定位仪在显示屏上的数字读数与根尖孔至文件尖端的距离的关系。

Distance from file tip to the major apical foramen in relation to the numeric meter reading on the display of three different electronic apex locators.

机构信息

Pulp Biology and Endodontics, Department of Restorative Sciences, Graduate School, Tokyo Medical and Dental University, Tokyo, Japan.

出版信息

Int Endod J. 2009 Dec;42(12):1065-70. doi: 10.1111/j.1365-2591.2009.01629.x.

Abstract

AIM

To establish and compare the relationship between the distance from the file tip to the apical foramen and the numeric meter reading on the display of three different electronic apex locators (EALs).

METHODOLOGY

A total of 12 extracted intact, straight, single-rooted human teeth with complete roots were used. The actual root canal length (AL) was determined after access preparation. For the electronic measurements with each EAL, silicon stops were fixed with auto-polymerizing resin to size 15 K-files at AL and 0.5, 1, 2, 3, 4 mm short of AL. The data was analysed by two-way anova and Tukey's honestly significant difference (HSD) test for multiple comparisons amongst EALs. Additionally, one-way anova and Tukey's HSD test were carried out for multiple comparisons amongst the measurements of each EAL.

RESULTS

There was a statistically significant difference amongst all EALs in indicating the position of file tips in relation to the major foramen (P < 0.05). The correlation between the meter reading and the position of the file tip from the apical foramen was statistically significant in the three EALs. There were significant differences amongst the measurements at distances from 0 to 2 mm in Justy III. In Dentaport, significant differences were found from 0 to 1 mm. However, the E-Magic Finder showed significant differences from 0 to 0.5 mm.

CONCLUSIONS

Justy III was more capable of displaying the intracanal position of the file tip from the major foramen in mm whilst advancing through the root canal during electronic measurements than the Dentaport and E-Magic Finder Deluxe.

摘要

目的

建立并比较三种不同电子根尖定位仪(EAL)的测杆尖端至根尖孔距离与数字读数之间的关系。

方法

共使用 12 颗完整、直、单根、无根的人离体牙。在进行牙髓腔预备后,确定实际根管长度(AL)。对于每个 EAL 的电子测量,用自凝树脂将硅止动器固定在尺寸为 15 K 锉上,使其位于 AL 和 AL 短 0.5、1、2、3、4 mm 处。采用双向方差分析和 Tukey 的 HSD 检验对三种 EAL 之间的电子测量数据进行多重比较。此外,还对每个 EAL 的测量结果进行了单向方差分析和 Tukey 的 HSD 检验。

结果

在指示测杆尖端相对于主孔位置方面,所有 EAL 之间均存在统计学差异(P < 0.05)。在三种 EAL 中,数字读数与测杆尖端从根尖孔位置之间存在统计学显著相关性。在 Justy III 中,从 0 到 2mm 的测量值之间存在显著差异。在 Dentaport 中,从 0 到 1mm 的测量值之间存在显著差异。然而,E-Magic Finder 从 0 到 0.5mm 的测量值之间存在显著差异。

结论

在电子测量过程中,Justy III 在显示测杆尖端从主孔进入根管内的位置方面比 Dentaport 和 E-Magic Finder Deluxe 更能以毫米为单位显示。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验