两种电子根尖定位仪的比较:一项体外研究。

A comparison between two electronic apex locators: an ex vivo investigation.

作者信息

Venturi M, Breschi L

机构信息

Department of Dental Sciences, Alma Mater Studiorum, University of Bologna, Bologna, Italy.

出版信息

Int Endod J. 2007 May;40(5):362-73. doi: 10.1111/j.1365-2591.2006.01229.x. Epub 2007 Feb 19.

Abstract

AIM

To compare ex vivo the performance of the Apex Finder and the Root ZX apex locators, with and without irrigant, in canals having different diameters.

METHODOLOGY

Sixty canals in 60 teeth were prepared using stainless steel hand files and 0.04 taper NiTi rotary instruments. During preparation the narrowest diameter of the canal was transported to the apical root surface. The canals were irrigated with RC-Prep and 5% NaOCl solution. Six groups were obtained, each with 10 canals having the same diameter of foramen, either 0.15, 0.20, 0.25, 0.40, 0.60 and 0.80 mm. A size 15 K-file was advanced into each canal until its tip was observed under x10 magnification to reach the foramen and the corresponding length recorded. The measurements were performed to an accuracy of 0.25 mm as a base unit of length. The teeth were then fixed to a plastic bar suspended over a glass container filled with 0.9% NaCl solution. Each apex locator was tested when the K-file was at the foramen, or 0.5, 1.0, 1.5 and 2.0 mm short; with the root apex immersed into the solution; with the canal dry or irrigated with NaCl. To evaluate the accuracy of both electronic apex locators (EALs) each electronically determined distance was compared with the actual length and the data analysed using the General Linear Model and the Student t-test.

RESULTS

Out of 2400 measurements 100 were electrically unstable, all with the Root ZX. In total, 521 measurements located the position of the file tip beyond the apex, in general, in high conductive conditions with the Root ZX and in low conductive conditions with the Apex Finder. No significant difference in terms of accuracy was found between the two EALs when the file tip was at the foramen (Root ZX mean +0.12 mm, SD 1.22 mm; Apex Finder mean +0.57 mm, SD 1.16 mm). Comparing all the measurements performed with the file tip within 2 mm of the foramen, in all the different conditions tested, the accuracy was affected (P<0.025) by diameter of the foramen, type of EAL, distance to the apex, and by several interactions.

CONCLUSIONS

Under the different ex vivo conditions both EALs provided accurate measurements when the file tip was at the foramen. The accuracy of the Apex Finder was negatively influenced by high conductive conditions, whilst the Root ZX provided inaccurate and unstable measurements mostly in low conductive conditions.

摘要

目的

在有不同直径的根管中,离体比较根尖定位仪和Root ZX根尖定位器在有冲洗液和无冲洗液情况下的性能。

方法

使用不锈钢手动锉和0.04锥度的镍钛旋转器械对60颗牙齿中的60个根管进行预备。预备过程中,将根管最窄直径处移至根尖牙根表面。用RC-Prep和5%次氯酸钠溶液冲洗根管。共得到六组,每组有10个根管,其根尖孔直径分别为0.15、0.20、0.25、0.40、0.60和0.80毫米。将一根15号K锉插入每个根管,直至在10倍放大倍数下观察到其尖端到达根尖孔,并记录相应长度。测量精度以0.25毫米作为长度基本单位。然后将牙齿固定在悬挂于装有0.9%氯化钠溶液的玻璃容器上方的塑料棒上。当K锉位于根尖孔处、或距根尖孔短0.5、1.0、1.5和2.0毫米时;根尖浸入溶液中时;根管干燥或用氯化钠冲洗时,对每个根尖定位器进行测试。为评估两种电子根尖定位器(EALs)的准确性,将每个电子测定的距离与实际长度进行比较,并使用一般线性模型和学生t检验对数据进行分析。

结果

在2400次测量中,有100次电不稳定,均使用Root ZX时出现。总体而言,在共521次测量中,锉尖位置超出根尖,通常在Root ZX的高传导条件下以及Apex Finder的低传导条件下出现。当锉尖位于根尖孔时,两种EALs在准确性方面未发现显著差异(Root ZX平均值+0.12毫米,标准差1.22毫米;Apex Finder平均值+0.57毫米,标准差1.16毫米)。比较在锉尖距根尖孔2毫米范围内进行的所有测量,在所有测试的不同条件下,准确性受根尖孔直径、EAL类型、距根尖距离以及几种相互作用的影响(P<0.025)。

结论

在不同的离体条件下,当锉尖位于根尖孔时,两种EALs均能提供准确测量。Apex Finder的准确性受到高传导条件的负面影响,而Root ZX大多在低传导条件下提供不准确和不稳定的测量结果。