Effect of stem modulus in a total hip arthroplasty model.

作者信息

Maistrelli G L, Fornasier V, Binnington A, McKenzie K, Sessa V, Harrington I

机构信息

University of Toronto, Ontario, Canada.

出版信息

J Bone Joint Surg Br. 1991 Jan;73(1):43-6. doi: 10.1302/0301-620X.73B1.1991773.

Abstract

The purpose of this study was to determine the biological effects of the elastic modulus of the femoral stem in canine hip arthroplasty. Cementless total hip arthroplasty was performed in 12 dogs, six had a low elastic modulus polyacetal resin stem and six had a high modulus stainless steel stem. The components were otherwise similar. At six and 12 months after operation, radiographic and histomorphometric analysis showed that those with steel implants had more cortical porosity than did the other group (p less than 0.01). We suggest that the elastic modulus of the implant is an important factor in controlling cortical bone resorption. A low modulus femoral prosthesis can significantly decrease bone resorption which might otherwise eventually lead to implant failure.

摘要

本研究的目的是确定犬髋关节置换术中股骨柄弹性模量的生物学效应。对12只犬实施了非骨水泥型全髋关节置换术,6只植入低弹性模量的聚缩醛树脂柄,6只植入高模量不锈钢柄。其他组件相似。术后6个月和12个月时,影像学和组织形态计量学分析显示,植入钢柄的犬皮质骨孔隙率高于另一组(p<0.01)。我们认为植入物的弹性模量是控制皮质骨吸收的一个重要因素。低模量股骨假体可显著减少骨吸收,否则最终可能导致植入物失败。