A study of experimental measurements of dosimetric parameters in HDR IR-192 source.

作者信息

Zhang Menglong, Zhang Liangan, Yuan Shuyu, Dai Guangfu

机构信息

Department of Health Physics, Institute of Radiation Medicine, Chinese Academy of Medical Sciences, Peking Union Medical College, Tianjin, China.

出版信息

Med Dosim. 2010 Winter;35(4):250-4. doi: 10.1016/j.meddos.2009.07.002. Epub 2009 Sep 17.

Abstract

Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them.

摘要

热释光剂量计已被用于使用改进的聚苯乙烯模体对广泛使用的Ir-192微型Selectron-HDR源进行剂量测定测量。已通过实验获得了径向剂量函数和各向异性函数,它们是TG-43U1协议中二维剂量率形式的主要参数。将测量结果与所报道的蒙特卡罗计算结果进行了比较,未发现它们之间存在差异。