Broda R, Dziel T, Muklanowicz A, Listkowska A, Pieńkowski Ł, Patocka A, Kołakowska E
Laboratory of Radioactivity Standards, Institute of Atomic Energy POLATOM Radioisotope Centre, Swierk, Poland.
Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1278-81. doi: 10.1016/j.apradiso.2009.11.017. Epub 2009 Nov 13.
The Radioisotope Centre (RC) has organized for the first time in Polish hospitals a comparison of activity measurements of (99)Tc(m) and (131)I. From the 58 Polish hospitals invited, 37 participated in the comparison. Each participant got about 4 mL of (99)Tc(m) eluate and two (131)I capsules, low- and high-activity. Sources were standardized in the RC before dispatch using a 4pi ionization chamber calibrated by standard solutions. Participants used their own radionuclide calibrators. The measurement results were evaluated using the E(n) criterion and the z'-score. From all 118 measurement results received from participants, 75 results (64%) lay within +/-5% of the RC value, and 99 results (84%) lay within +/-10.0%.