Institute of Anaesthesiology, University Hospital Zurich, Zurich, Switzerland.
Eur J Anaesthesiol. 2010 Mar;27(3):289-94. doi: 10.1097/EJA.0b013e3283352201.
In-vitro performance of the PediaSat system for continuous monitoring of central venous oxygen saturation by spectrophotometry has been evaluated.
PediaSat continuous fibre-optic oximetry catheters were inserted in a black testing chamber, connected with an extracorporeal circuit and filled with human whole blood. Oxygen inflow into the cardiopulmonary bypass system was varied, and the testing chamber was perfused with blood flow of 1000 ml min(-1). Oxygen saturation values measured by PediaSat (S PediaSat O2) were compared with cooximetry (S CO-OX O2) values from simultaneously taken blood samples by Bland-Altman and simple regression analyses.
Fifty data pairs were obtained. S PediaSat O2 and S CO-OX O2 values ranged between 28-98 and 24.9-99.5%, respectively. Correlation between S PediSat O2 and S CO-OX O2 was high with an r2 value equal to 0.96 (P < 0.0001). Overall, S PediaSat O2 only slightly overestimated S CO-OX O2 (mean bias +2.9%), and limits of agreement (+/-2 SD of bias) were acceptable (-6.8/+12.6%). Sensitivity and specificity of the first differences of S PediaSat O2 and S CO-OX O2 were 1.0 and 0.92, respectively. Subgroup analysis of S CO-OX O2 values below 70% resulted in an overestimation by S PediaSat O2, with a mean bias of +5.2% and limits of agreement of -4.7 and +15.1%.
The current version of the PediaSat system does not reliably reflect S CO-OX O2 values below 70%, but it seems to be a useful tool providing an accurate trend of continuous central venous oxygen saturation.
评估了 PediaSat 系统通过分光光度法连续监测中心静脉血氧饱和度的体外性能。
将 PediaSat 连续光纤血氧饱和度导管插入黑色测试室,与体外循环连接,并充满人体全血。改变心肺旁路系统中的氧气流入量,以 1000ml/min 的血流量灌注测试室。通过 Bland-Altman 分析和简单回归分析,将 PediaSat 测量的血氧饱和度值(S PediaSat O2)与同时采集的血液样本的 cooximetry(S CO-OX O2)值进行比较。
获得了 50 对数据。S PediaSat O2 和 S CO-OX O2 值分别在 28-98 和 24.9-99.5%之间。S PediSat O2 与 S CO-OX O2 之间的相关性很高,r2 值等于 0.96(P<0.0001)。总体而言,S PediaSat O2 仅略微高估了 S CO-OX O2(平均偏差+2.9%),且一致性界限(偏差+/-2 SD)可接受(-6.8/+12.6%)。S PediaSat O2 和 S CO-OX O2 的第一差异的灵敏度和特异性分别为 1.0 和 0.92。S CO-OX O2 值低于 70%的亚组分析导致 S PediaSat O2 高估,平均偏差为+5.2%,一致性界限为-4.7 和+15.1%。
当前版本的 PediaSat 系统不能可靠地反映 S CO-OX O2 值低于 70%的情况,但它似乎是一种有用的工具,可提供连续中心静脉血氧饱和度的准确趋势。