Suppr超能文献

干电极心电图记录中的放大器输入阻抗。

Amplifier input impedance in dry electrode ECG recording.

作者信息

Assambo Cedric, Burke Martin J

机构信息

Dept. of Electronic & Electrical Engineering, Trinity College Dublin, Dublin 2, Rep. of Ireland.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2009;2009:1774-7. doi: 10.1109/IEMBS.2009.5333398.

Abstract

This paper presents a novel approach for designing the front-end of instrumentation amplifiers for use in dry electrode recording of the human electrocardiogram (ECG). The method relies on information provided by the characterization of the skin-electrode interface and the analysis of low frequency ECG criteria defined by international standards. Marginal measurements of capacitive elements of the skin-electrode interface as small as 0.01 microF, suggest values of input impedance in the order of 1.3 GOmega. However, results in 99% of the data analyzed indicate that a recording amplifier providing an input impedance of 500 MOmega should ensure clear signal sensing without distortion.

摘要

本文提出了一种用于设计仪器放大器前端的新颖方法,该放大器用于人体心电图(ECG)的干电极记录。该方法依赖于皮肤-电极界面表征所提供的信息以及对国际标准定义的低频ECG标准的分析。皮肤-电极界面电容元件的微小测量值低至0.01微法,表明输入阻抗值约为1.3吉欧。然而,对99%的分析数据的结果表明,提供500兆欧输入阻抗的记录放大器应能确保清晰的信号传感且无失真。