Dozio Roberta, Baba Adeshina, Assambo Cedric, Burke Martin J
Dept. of Electronic & Electrical Engineering, Trinity College Dublin, College Green, Dublin 2, Rep. of Ireland.
Annu Int Conf IEEE Eng Med Biol Soc. 2007;2007:5001-4. doi: 10.1109/IEMBS.2007.4353463.
This paper reports the measurement of the properties of dry or pasteless conductive electrodes to be used for long-term recording of the human electrocardiogram (ECG). Knowledge of these properties is essential for the correct design of the input stage of associated recording amplifiers. Measurements were made on three commercially available conductive carbon based electrodes at pressures of 5 mmHg and 20 mmHg, located on the lower abdomen of the body on three subjects having different skin types. Parameter values were fitted to a two-time-constant based model of the electrode using data measured over a period of 10s. Values of resistance, ranging from 40kOmega to 1590kOmega and of capacitance ranging from 0.05 microF to 38 microF were obtained for the components, while the values of the time-constants varied from 0.07 s to 3.9s.
本文报道了用于长期记录人体心电图(ECG)的干式或无膏导电电极特性的测量。了解这些特性对于相关记录放大器输入级的正确设计至关重要。在三名具有不同皮肤类型的受试者的下腹部,对三种市售的碳基导电电极在5 mmHg和20 mmHg的压力下进行了测量。使用在10秒内测量的数据,将参数值拟合到基于双时间常数的电极模型。这些组件的电阻值范围为40kΩ至1590kΩ,电容值范围为0.05μF至38μF,而时间常数的值则从0.07 s变化到3.9s。