Central Isotope and Clinical Laboratory, Semmelweis University, Hungary.
Clin Chim Acta. 2010 Mar;411(5-6):364-7. doi: 10.1016/j.cca.2009.11.033. Epub 2009 Dec 5.
Midnight salivary cortisol (MSC) concentration has been considered as a sensitive marker of overt hypercortisolism. Because no studies on commercially available automated, non-isotopic MSC assays have been reported, we determined and compared the diagnostic performance of an automated electrochemiluminescent immunoassay (ECLIA, Elecsys E170) and an in-house radioimmunoassay (RIA) for MSC measurement.
The study involved 126 consecutive patients referred for evaluation because of symptoms of Cushing's syndrome, obesity, or the presence of incidentally discovered adrenal adenoma. Using detailed clinical, hormonal and radiological evaluation overt endogenous hypercortisolism was confirmed in 9 patients and was excluded in 117 patients.
ROC analysis indicated that the best performance of MSC was obtained at cutoff value of 0.35 microg/dl using ECLIA (sensitivity, 100%; specificity, 88%) and 0.29 microg/dl (sensitivity, 100%; specificity, 71%) using RIA. When the findings were compared to those obtained from low dose dexamethasone test, both ECLIA and RIA of MSC showed a better diagnostic performance.
MSC measurement is useful for the diagnosis of overt hypercortisolism but the cutoff value is highly dependent on the method used. We recommend the use of automated ECLIA for MSC assay, and we propose further studies on other automated immunoassay analyzers potentially suitable for MSC measurements.
午夜唾液皮质醇(MSC)浓度被认为是显性皮质醇增多症的敏感标志物。由于没有关于市售自动化、非同位素 MSC 检测的研究报告,我们确定并比较了用于 MSC 测量的自动化电化学发光免疫分析(ECLIA,Elecsys E170)和内部放射免疫分析(RIA)的诊断性能。
该研究涉及 126 例连续患者,这些患者因库欣综合征、肥胖或偶然发现的肾上腺腺瘤的症状而被转介进行评估。使用详细的临床、激素和影像学评估,确认 9 例患者存在显性内源性皮质醇增多症,并排除 117 例患者。
ROC 分析表明,ECLIA 的最佳 MSC 表现为 0.35 μg/dl 截断值(敏感性为 100%,特异性为 88%),RIA 的最佳 MSC 表现为 0.29 μg/dl(敏感性为 100%,特异性为 71%)。当将这些发现与小剂量地塞米松试验的结果进行比较时,ECLIA 和 RIA 的 MSC 均表现出更好的诊断性能。
MSC 测量对于显性皮质醇增多症的诊断很有用,但截断值高度依赖于所用方法。我们建议使用自动化 ECLIA 进行 MSC 检测,并建议对其他潜在适合 MSC 测量的自动化免疫分析分析仪进行进一步研究。