Assessment of accuracy of PET utilizing a 3-D phantom to simulate the activity distribution of [18F]fluorodeoxyglucose uptake in the human brain.

作者信息

Hoffman E J, Cutler P D, Guerrero T M, Digby W M, Mazziotta J C

机构信息

Division of Nuclear Medicine and Biophysics, UCLA School of Medicine 90024.

出版信息

J Cereb Blood Flow Metab. 1991 Mar;11(2):A17-25. doi: 10.1038/jcbfm.1991.32.

Abstract

A three-dimensional brain phantom has been developed to simulate the activity distributions found in human brain studies currently employed in positron emission tomography (PET). The phantom has a single contiguous chamber and utilizes thin layers of lucite to provide apparent relative concentrations of 5, 1, and 0 for gray matter, white matter, and CSF structures, respectively. The phantom and an ideal image set were created from the same set of data. Thus, the user has a basis for comparing measured images with an ideal set that allows a quantitative evaluation of errors in PET studies with an activity distribution similar to that found in patients. The phantom was employed in a study of the effect of deadtime and scatter on accuracy in quantitation on a current PET system. Deadtime correction factors were found to be significant (1.1-2.5) at count rates found in clinical studies. Deadtime correction techniques were found to be accurate to within 5%. Scatter in emission and attenuation correction data consistently caused 5-15% errors in quantitation, whereas correction for scatter in both types of data reduced errors in accuracy to less than 5%.