Kummer Frederick J, Pereira Gavin, Schachter Aaron K, Jaffe William L
NYU School of Medicine, New York, New York, USA.
Bull NYU Hosp Jt Dis. 2009;67(4):341-6.
Previous studies have suggested that femoral component positioning in resurfacing arthroplasty may affect strains in the femoral neck that could lead to decreased implant longevity. A strain gaged, Sawbones model was used to determine the femoral neck strains for a variety of resurfacing head translations and angulations. We found that head positions affected strain distributions, most positions leading to increased neck strains, often over 100%, with the exception being a varus head position where the superior neck strains decreased over 50%. Although the clinical meaning of these findings is unclear, it could be of concern for stress-shielding or fatigue fracture of the femoral neck.