Suppr超能文献

骨盆测量磁共振成像,观察者间和观察者内变异分析。

MR pelvimetry measurements, analysis of inter- and intra-observer variation.

机构信息

Department of Obstetrics and Gynecology, Kuopio University Hospital, Finland.

出版信息

Eur J Radiol. 2010 Aug;75(2):e56-61. doi: 10.1016/j.ejrad.2009.11.018. Epub 2009 Dec 14.

Abstract

OBJECTIVE

To set reference standards and then to evaluate the measurement variation in magnetic resonance (MR) pelvimetry between observations and observers.

METHODS

The study was carried out retrospectively using 100 MR pelvimetry examinations performed in North Carelian Central Hospital between September 2006 and January 2008. Pelvimetric parameters of pelvic inlet and outlet were measured four times to determine the standard reference for each measurement and then intra- and inter-observer variations were compared.

RESULTS

The accuracy of MR pelvimetry, defined as a deviation of <5mm from the reference standard, was better in pelvic inlet measurements than in the corresponding outlet measurements (95-99% vs. 86-89%). Intra-observer variation was acceptable in all of the measurements with intraclass correlation coefficient (ICC) in the range 0.956-0.981 in all parameters. Inter-observer variation was higher than intra-observer variation. The largest variation of measurements was in pelvic outlet parameters between observers with the ICC in the range of 0.710-0.813.

CONCLUSION

MR pelvimetry measurement should be conducted in a centralized location to decrease observer-related variation. Clinicians should be aware that millimeter differences are not reliable in MR pelvimetry and therefore the use of millimeter accurate limits are not recommended in obstetric decision making.

摘要

目的

制定参考标准,然后评估磁共振(MR)骨盆测量在观察者之间的测量变化。

方法

该研究采用回顾性研究方法,对 2006 年 9 月至 2008 年 1 月在北卡累利阿中央医院进行的 100 次 MR 骨盆测量进行研究。对骨盆入口和出口的骨盆参数进行了 4 次测量,以确定每个测量的标准参考值,然后比较了观察者内和观察者间的差异。

结果

以参考标准的偏差<5mm 为定义的磁共振骨盆测量的准确性,在骨盆入口测量中优于相应的出口测量(95-99%比 86-89%)。所有参数的观察者内变异均在可接受范围内,组内相关系数(ICC)范围为 0.956-0.981。观察者间的变异高于观察者内的变异。观察者之间的骨盆出口参数测量差异最大,ICC 范围为 0.710-0.813。

结论

MR 骨盆测量应在集中的位置进行,以减少观察者相关的变异。临床医生应意识到,在磁共振骨盆测量中,毫米差异不可靠,因此不建议在产科决策中使用毫米精确的限制。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验