Suppr超能文献

采用碲化镉锌探测器技术的核心肌灌注成像:扫描时间优化方案。

Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction.

机构信息

Cardiac Imaging, University Hospital Zurich, CH-8091 Zurich, Switzerland.

出版信息

J Nucl Med. 2010 Jan;51(1):46-51. doi: 10.2967/jnumed.109.065532. Epub 2009 Dec 15.

Abstract

UNLABELLED

We aimed at establishing the optimal scan time for nuclear myocardial perfusion imaging (MPI) on an ultrafast cardiac gamma-camera using a novel cadmium-zinc-telluride (CZT) solid-state detector technology.

METHODS

Twenty patients (17 male; BMI range, 21.7-35.5 kg/m(2)) underwent 1-d (99m)Tc-tetrofosmin adenosine stress and rest MPI protocols, each with a 15-min acquisition on a standard dual-detector SPECT camera. All scans were immediately repeated on an ultrafast CZT camera over a 6-min acquisition time and reconstructed from list-mode raw data to obtain scan durations of 1 min, 2 min, etc., up to a maximum of 6 min. For each of the scan durations, the segmental tracer uptake value (percentage of maximum myocardial uptake) from the CZT camera was compared by intraclass correlation with standard SPECT camera data using a 20-segment model, and clinical agreement was assessed per coronary territory. Scan durations above which no further relevant improvement in uptake correlation was found were defined as minimal required scan times, for which Bland-Altman limits of agreement were calculated.

RESULTS

Minimal required scan times were 3 min for low dose (r = 0.81; P < 0.001; Bland-Altman, -11.4% to 12.2%) and 2 min for high dose (r = 0.80; P < 0.001; Bland-Altman, -7.6% to 12.9%), yielding a clinical agreement of 95% and 97%, respectively.

CONCLUSION

We have established the minimal scan time for a CZT solid-state detector system, which allows 1-d stress/rest MPI with a substantially reduced acquisition time resulting in excellent agreement with regard to uptake and clinical findings, compared with MPI from a standard dual-head SPECT gamma-camera.

摘要

目的

本研究旨在利用新型碲化镉锌(CZT)固态探测器技术,为超快心脏伽马相机确定核心肌灌注成像(MPI)的最佳扫描时间。

方法

20 例患者(男 17 例;BMI 范围为 21.7-35.5kg/m²)行 1 天(99m)Tc-甲氧基异丁基异腈腺苷负荷和静息 MPI 方案,每个方案均在标准双探头 SPECT 相机上进行 15 分钟采集。所有扫描立即在超快 CZT 相机上重复进行,采集时间为 6 分钟,并从列表模式原始数据重建,获得 1 分钟、2 分钟等的扫描时间,最长达 6 分钟。对于每个扫描时间,使用 20 节段模型,通过内类相关,将 CZT 相机的节段示踪剂摄取值(最大心肌摄取百分比)与标准 SPECT 相机数据进行比较,并根据每个冠状动脉节段评估临床一致性。将无进一步摄取相关性显著改善的扫描时间定义为最短必需扫描时间,计算其 Bland-Altman 协议界限。

结果

低剂量时最短必需扫描时间为 3 分钟(r = 0.81;P < 0.001;Bland-Altman,-11.4%至 12.2%),高剂量时最短必需扫描时间为 2 分钟(r = 0.80;P < 0.001;Bland-Altman,-7.6%至 12.9%),分别获得 95%和 97%的临床一致性。

结论

我们确定了 CZT 固态探测器系统的最短扫描时间,与标准双探头 SPECT 伽玛相机相比,该系统可在大幅缩短采集时间的情况下进行 1 天的负荷/静息 MPI,摄取和临床结果具有极好的一致性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验