Tang Min, Kriatselis Charalampos, Nedios Sotiris, Ye Gang, Roser Mathias, Fleck Eckard, Gerds-Li Jin-Hong
Department of Internal Medicine/Cardiology, German Heart Institute Berlin, Berlin, Germany.
J Cardiovasc Electrophysiol. 2010 Jun 1;21(6):626-31. doi: 10.1111/j.1540-8167.2009.01668.x. Epub 2009 Dec 15.
The study was designed to evaluate the feasibility and efficacy of a simplified cryoballoon technique in which a microcircular catheter was introduced into the central lumen of a cryoballoon catheter for the purpose of recording pulmonary vein (PV) potentials during ablation procedures and without interchanging catheters.
A total of 23 consecutive patients with paroxysmal atrial fibrillation (AF) were enrolled. A single transseptal puncture was made and a cryoballoon catheter was inserted into the left atrium. A 6-pole mapping catheter with a 0.035-inch shaft diameter was introduced into the PV through the central lumen of the cryoballoon catheter. In addition to the function as a recording device, the mapping catheter was also used as a "guide-wire" during the procedure. A total of 84 PVs (84/92, 91.3%) were completely isolated using this novel cryoballoon technique. In 43 of the 84 veins (51.2%), isolation was observed in real time during the cryoablation; in the remaining 41 veins (48.8%), isolation was confirmed immediately post ablation attempt with the mapping catheter. Procedure time was 152.7 +/- 54.9 minutes, and fluoroscopy time was 33.2 +/- 17.3 minutes. At follow-up (7.4 months, range 2-18 months), 17 (73.9%) patients were free from AF. There was 1 occurrence of phrenic nerve palsy during ablation of a right superior PV, which fully resolved after 1 month.
The use of a cryoballoon catheter equipped with a 6-pole micromapping catheter inserted through its central lumen for the purpose of mapping and ablation during PV isolation procedures is both feasible and effective.
本研究旨在评估一种简化冷冻球囊技术的可行性和有效性,该技术将微圆形导管引入冷冻球囊导管的中心腔,以便在消融过程中记录肺静脉(PV)电位,且无需更换导管。
连续纳入23例阵发性心房颤动(AF)患者。进行单次经房间隔穿刺,将冷冻球囊导管插入左心房。通过冷冻球囊导管的中心腔将一根轴径为0.035英寸的6极标测导管引入肺静脉。除作为记录装置外,标测导管在手术过程中还用作“导丝”。使用这种新型冷冻球囊技术共成功隔离了84条肺静脉(84/92,91.3%)。在84条肺静脉中的43条(51.2%)中,在冷冻消融过程中实时观察到隔离;在其余41条肺静脉(48.8%)中,消融尝试后立即用标测导管确认隔离。手术时间为152.7±54.9分钟,透视时间为33.2±17.3分钟。随访(7.4个月,范围2 - 18个月)时,17例(73.9%)患者无房颤发作。在右上肺静脉消融过程中发生1例膈神经麻痹,1个月后完全恢复。
在肺静脉隔离手术中,使用通过其中心腔插入6极微标测导管的冷冻球囊导管进行标测和消融是可行且有效的。