A retrospective evaluation of the ability of computer-aided detection (CAD) ability to identify breast carcinoma in standard mammographic projections.
Forty-five biopsy proven lesions in 44 patients imaged digitally with CAD applied at examination were reviewed. Forty-four screening BIRADS category 1 digital mammography examinations were randomly identified to serve as a comparative normal/control population. Data included patient age; BIRADS breast density; lesion type, size, and visibility; number, type, and location of CAD marks per image; CAD ability to mark lesions; needle core and surgical pathologic correlation.
The CAD lesion/case sensitivity of 87% (n = 39), image sensitivity of 69% (n = 31) for mediolateral oblique view and 78% (n = 35) for the craniocaudal view was found. The average false positive rate in 44 normal screening cases was 2.0 (range 1-8). The 2.0 figure is based on 88 reported false positive CAD marks in 44 normal screening exams: 98% (n = 44) lesions proceeded to excision; initial pathology upgraded at surgical excision from in situ to invasive disease in 24% (n = 9) lesions.
CAD demonstrated potential to detect mammographically visible cancers in standard projections for all lesion types.
回顾性评估计算机辅助检测(CAD)在标准乳腺 X 线摄影投影中识别乳腺癌的能力。
对 44 例经 CAD 辅助检测并经数字成像证实有病变的患者的 45 个活检证实的病变进行回顾性研究。随机选取 44 例乳腺 X 线摄影 BI-RADS 类别为 1 的筛查病例作为正常/对照人群。数据包括患者年龄;乳腺密度;病变类型、大小和可见性;每个图像的 CAD 标记数量、类型和位置;CAD 标记病变的能力;针芯和手术病理相关性。
CAD 对病变/病例的检出率为 87%(n=39),对斜位(mediolateral oblique view)的图像检出率为 69%(n=31),对头尾位(craniocaudal view)的图像检出率为 78%(n=35)。44 例正常筛查病例的平均假阳性率为 2.0(范围 1-8)。这个 2.0 的数字是基于 44 例正常筛查检查中报告的 88 个假阳性 CAD 标记:98%(n=44)的病变进行了切除;初始病理在手术切除时从原位癌升级为浸润性癌的比例为 24%(n=9)。
CAD 显示出在标准投影中检测所有类型病变的乳腺 X 线摄影可见癌症的潜力。