Computer-aided detection in digital mammography: comparison of craniocaudal, mediolateral oblique, and mediolateral views.

作者信息

Kim Seung Ja, Moon Woo Kyung, Cho Nariya, Cha Joo Hee, Kim Sun Mi, Im Jung-Gi

机构信息

Department of Radiology and Clinical Research Institute, Seoul National University Hospital and the Institute of Radiation Medicine, Seoul National University Medical Research Center, 28, Yongon-dong, Chongno-gu, Seoul, 100-744, Korea.

出版信息

Radiology. 2006 Dec;241(3):695-701. doi: 10.1148/radiol.2413051145.

Abstract

PURPOSE

To retrospectively compare the sensitivity of a computer-aided detection (CAD) system for depicting breast cancer in three digital mammographic views.

MATERIALS AND METHODS

This study was conducted with institutional review board approval; informed consent was waived. A commercially available CAD system was applied to the craniocaudal, mediolateral oblique, and mediolateral digital mammographic views of 83 women (mean age, 48 years; range, 30-66 years) with 83 histologically proved breast cancers. Findings were 59 masses and 41 microcalcifications (17 lesions showed both findings; 42 lesions, mass only; and 24 lesions, microcalcification only). The paired t test was used to analyze sensitivity of the CAD system for the detection of cancer in these three mammographic views and in combinations of the views.

RESULTS

The sensitivities of the CAD system were 92% (76 of 83) in the craniocaudal view, 83% (69 of 83) in the mediolateral oblique view, and 86% (71 of 83) in the mediolateral view; the differences were not significant (P = .07-.62). Sensitivity increased to 96% (80 of 83) in the craniocaudal plus mediolateral oblique views and to 99% (82 of 83) in the craniocaudal plus mediolateral oblique plus mediolateral views. For masses, the sensitivity of the CAD system was 76% (45 of 59) in the craniocaudal view and 75% (44 of 59) in the mediolateral oblique view and increased to 93% (55 of 59) when mediolateral oblique and craniocaudal views were combined (P < .001). For microcalcifications, sensitivity was 98% (40 of 41) in the craniocaudal view and 95% (39 of 41) in the mediolateral oblique view, and this increased to 100% (41 of 41) when the mediolateral oblique and craniocaudal views were combined (P = .31).

CONCLUSION

The sensitivities of the CAD system were not significantly different among these three digital mammographic views. Sensitivity for depicting masses was significantly increased (P < .001) when the craniocaudal view was added to the mediolateral oblique view.

摘要

目的

回顾性比较计算机辅助检测(CAD)系统在三种数字化乳腺钼靶视图中检测乳腺癌的敏感性。

材料与方法

本研究经机构审查委员会批准进行;免除知情同意。将一种商用CAD系统应用于83例经组织学证实患有乳腺癌的女性(平均年龄48岁;范围30 - 66岁)的头尾位、内外斜位和内外侧数字化乳腺钼靶视图。发现59个肿块和41处微钙化(17处病变同时显示两种表现;42处病变仅有肿块;24处病变仅有微钙化)。采用配对t检验分析CAD系统在这三种乳腺钼靶视图及视图组合中检测癌症的敏感性。

结果

CAD系统在头尾位视图中的敏感性为92%(83例中的76例),在内外斜位视图中为83%(83例中的69例),在内外侧视图中为86%(83例中的71例);差异无统计学意义(P = 0.07 - 0.62)。在头尾位加内外斜位视图中敏感性增至96%(83例中的80例),在头尾位加内外斜位加内外侧视图中增至99%(83例中的82例)。对于肿块,CAD系统在头尾位视图中的敏感性为76%(59例中的45例),在内外斜位视图中为75%(59例中的44例),当内外斜位视图与头尾位视图组合时增至93%(59例中的55例)(P < 0.001)。对于微钙化,在头尾位视图中的敏感性为98%(41例中的40例),在内外斜位视图中为95%(41例中的39例),当内外斜位视图与头尾位视图组合时增至100%(41例中的41例)(P = 0.31)。

结论

CAD系统在这三种数字化乳腺钼靶视图中的敏感性无显著差异。当在内外斜位视图基础上增加头尾位视图时,检测肿块的敏感性显著提高(P < 0.001)。