Irradiation of 4''x4'' NaI(Tl) detector by the 14 MeV neutrons.

作者信息

Sudac D, Valkovic V

机构信息

Rudjer Boskovic Institute, Zagreb, Croatia.

出版信息

Appl Radiat Isot. 2010 Apr-May;68(4-5):896-900. doi: 10.1016/j.apradiso.2009.12.014. Epub 2009 Dec 16.

Abstract

Within the EURopean Illicit TRAfficking Countermeasures Kit (EURITRACK) project, a new Tagged Neutron Inspection System (TNIS) has been developed and installed in the Port of Rijeka in Croatia. The system was based on the examination of sea containers with the 14 MeV neutron beam. During the operation the characteristic gamma rays were produced and measured by several 5''x5''x10'' NaI(Tl) detectors. During this procedure some of the detectors were exposed to an intensive neutron beam radiation. It was necessary to check for possible radiation damage of the NaI(Tl) scintillator during the gamma detector selection phase of the project. The 4''x4'' NaI(Tl) detector was exposed to 14 MeV neutrons for 20 h. From the presented results on energy resolution and activation measurements it could be concluded that there are no significant differences in energy resolution before and after the irradiation by 4.7x10(11) of 14 MeV neutrons. The only problem could be the high level of medium and long term induced activity in the energy region below 2 MeV.