手动和电动刚度计在定量牵张成骨中骨再生方面的比较。

Comparison of a manual and motorized stiffness meter to quantify bone regeneration in distraction osteogenesis.

作者信息

Thorey Fritz, Floerkemeier Thilo, Wellmann Matthias, Windhagen Henning

机构信息

Department of Orthopaedic Surgery, Hannover Medical School, Germany.

出版信息

Technol Health Care. 2009;17(5-6):369-75. doi: 10.3233/THC-2009-0556.

Abstract

To assess bone healing and investigate the influence of different pharmaceutics (e.g. growth factors) on bone stiffness and strength in-vivo, new quantitative methods are necessary. Therefore, a new manual and motorized stiffness meter to quantify bone regeneration in a model of distraction osteogenesis were compared. The design, equipment, and improvements of the measurement devices are described. Furthermore, their difference in precision and accuracy in comparison to tests from a material testing system, used as "gold standard", were evaluated. Both devices were able to assess regenerate stiffness: the accuracy ranged between +/- 9% for the manual and +/-5% for the motorized version for stiffness data over 0.1 Nm/ degrees; precision between +/- 3.8% for the manual and +/- 3.2% for the motorized device. In summary, the two stiffness measurement devices described in this study have the power to monitor the beginning of bone healing and therefore predict the load bearing capacity of regenerating bone. The motorized version showed advantages over the manual device when investigating and monitoring the stiffness of bone during a consolidation period: (1) better accuracy in both stiffness below and above 0.1 Nm/ degrees, (2) a better precision in the stiffness range of interest, (3) easier handling, and (4) standardisation of the measurement process using the stepper motor and definition of the maximums of torque, angulation and rotation speed.

摘要

为了评估骨愈合情况,并研究不同药物制剂(如生长因子)对体内骨硬度和强度的影响,需要新的定量方法。因此,比较了一种新型手动和电动硬度计在牵张成骨模型中量化骨再生的情况。描述了测量装置的设计、设备和改进之处。此外,还评估了它们与作为“金标准”的材料测试系统测试相比在精度和准确性方面的差异。两种装置都能够评估再生骨的硬度:对于超过0.1 Nm/度的硬度数据,手动装置的准确度在±9%之间,电动装置在±5%之间;手动装置的精度在±3.8%之间,电动装置在±3.2%之间。总之,本研究中描述的两种硬度测量装置有能力监测骨愈合的开始,从而预测再生骨的承载能力。在研究和监测骨愈合期的骨硬度时,电动版本比手动装置具有优势:(1)在低于和高于0.1 Nm/度的硬度方面都有更好的准确度,(2)在感兴趣的硬度范围内有更好的精度,(3)操作更简便,(4)使用步进电机使测量过程标准化,并定义了扭矩、角度和转速的最大值。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验