Stevenson R W, Ansell R O
Department of Ophthalmic Optics, Glasgow College, Scotland.
CLAO J. 1991 Jan;17(1):36-40.
A polarographic method of oxygen determination across contact lenses has been developed. A custom designed polarographic cell, with a platinum cathode and a silver anode was used to measure a series of specially prepared contact lens samples made from four gas permeable contact lens materials currently commercially available. On a series of measurements the polarographic current was shown to be highly repeatable without the need for a wet membrane between the lens and the cathode. A layer of solution was adequate. The greatest variability was found with the thinnest sample of the highest oxygen permeable material. The results were plotted using the resistance (L/Dk) of the readings against sample thickness. Three of the four examples showed a curvilinear function and the other a linear relationship relative to sample thickness. From these graphs the reciprocal of the slope of the curve at zero thickness gave the measured oxygen permeability of the material. The significance of this system towards setting up a standard for the measurement of oxygen permeability of contact lens materials is discussed.
已开发出一种通过隐形眼镜测定氧气的极谱法。使用定制设计的极谱池,其具有铂阴极和银阳极,来测量一系列由目前市场上可买到的四种透气隐形眼镜材料制成的特制隐形眼镜样品。在一系列测量中,极谱电流显示出高度可重复性,无需在镜片和阴极之间设置湿膜。一层溶液就足够了。在最高透氧材料的最薄样品中发现了最大的变异性。使用读数的电阻(L/Dk)相对于样品厚度绘制结果。四个示例中的三个显示出曲线函数,另一个相对于样品厚度呈线性关系。从这些图表中,零厚度处曲线斜率的倒数给出了材料的测量透氧率。讨论了该系统对于建立隐形眼镜材料透氧率测量标准的意义。