肩胛盂松质骨的力学性能。
Mechanical properties of glenoid cancellous bone.
作者信息
Kalouche Ibrahim, Crépin Jérôme, Abdelmoumen Samir, Mitton David, Guillot Geneviève, Gagey Olivier
机构信息
Bicetre University Hospital, AP-HP Paris, F-94270, France.
出版信息
Clin Biomech (Bristol). 2010 May;25(4):292-8. doi: 10.1016/j.clinbiomech.2009.12.009. Epub 2010 Jan 18.
BACKGROUND
Loosening of the glenoid component in total shoulder arthroplasty is the main late complication of this procedure; it may be assumed that it is highly dependent on the quality of the glenoid cancellous bone. Very little is known about the mechanical properties of this cancellous bone. The aim of this study was to determine these properties (Young's modulus and strength) as well as bone density in different parts of the glenoid cancellous bone to assess their variations.
METHODS
Eleven scapulas were obtained from six fresh-frozen, unembalmed human cadavers. Eighty-two cubic cancellous bone specimens were extracted and tested using a uniaxial compression test; then the specimens were defatted and correlations with bone density were determined.
FINDINGS
The study showed significant differences in the mechanical properties with anatomic location and directions of loading. Young's modulus and strength were found to be significantly higher at the posterior part of the glenoid with the weakest properties at the antero-inferior part. Cancellous bone was found to be anisotropic with higher mechanical properties in the latero-medial direction perpendicular to the articular surface of the glenoid. The apparent density was on average equal to 0.29 g/cm(3) with the higher values at the posterior and superior part of the glenoid. Good correlation between apparent density and elastic modulus was found only in the sagittal planes but not in the coronal and axial plane.
INTERPRETATION
The mechanical properties determined in this study showed the anisotropy of the glenoid cancellous bone; values of these properties could provide input data for finite element method analyses in shoulder prosthesis designs.
背景
全肩关节置换术中肩胛盂假体松动是该手术的主要晚期并发症;可以认为,这高度依赖于肩胛盂松质骨的质量。关于这种松质骨的力学性能知之甚少。本研究的目的是确定这些性能(杨氏模量和强度)以及肩胛盂松质骨不同部位的骨密度,以评估其变化。
方法
从6具新鲜冷冻、未防腐处理的人类尸体上获取11块肩胛骨。提取82个立方松质骨标本并使用单轴压缩试验进行测试;然后将标本脱脂并确定与骨密度的相关性。
结果
研究表明,力学性能在解剖位置和加载方向上存在显著差异。发现肩胛盂后部的杨氏模量和强度显著更高,而在前下部分性能最弱。松质骨表现为各向异性,在垂直于肩胛盂关节面的内外侧方向上具有更高的力学性能。表观密度平均等于0.29 g/cm³,在肩胛盂的后部和上部值更高。仅在矢状面发现表观密度与弹性模量之间有良好的相关性,而在冠状面和轴面则没有。
解读
本研究确定的力学性能显示了肩胛盂松质骨的各向异性;这些性能的值可为肩部假体设计中的有限元分析提供输入数据。