Lim E S, Apple D J, Tsai J C, Morgan R C, Wasserman D, Assia E I
Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston.
Ophthalmology. 1991 Feb;98(2):243-6. doi: 10.1016/s0161-6420(91)32310-8.
A survey of 1204 closed-loop anterior chamber intraocular lenses (AC-IOLs) and 310 open-loop AC-IOLs accessioned between November 1982 and January 1990 was conducted at the Center for Intraocular Lens Research. An analysis of complication rates was done after normalization of data with respect to market share totals. The results establish that an unacceptable complication rate is associated with the closed-loop design when compared with either the tripod or quadripod lens styles. Furthermore, the closed-loop designs, while comprising an estimated 45% of the total number of AC-IOLs estimated to be implanted in the United States (n = 674,000), were responsible for 80% of the AC-IOLs explanted after complications and accessioned at the authors' center. A rethinking of the extreme condemnation of all anterior chamber IOLs that has surfaced in recent years is warranted. This is particularly true with respect to indications for use of sutured posterior chamber (PC) IOLs as well as with regard to possible use of open-loop AC-IOLs in less-industrialized nations.
眼内晶状体研究中心对1982年11月至1990年1月间登记入册的1204枚闭环式前房人工晶状体(AC-IOL)和310枚开环式AC-IOL进行了一项调查。在对数据按市场份额总数进行标准化处理后,对并发症发生率进行了分析。结果表明,与三脚架式或四脚架式晶状体类型相比,闭环设计的并发症发生率令人无法接受。此外,闭环设计虽然估计占美国预计植入的AC-IOL总数的45%(n = 674,000),但在并发症发生后被取出并在作者所在中心登记入册的AC-IOL中,有80%是闭环设计的。有必要重新思考近年来对所有前房人工晶状体的极端谴责。对于缝合式后房(PC)人工晶状体的使用指征以及在工业化程度较低的国家中开环式AC-IOL的可能使用情况,尤其如此。