Laboratoire des IMRCP, Université de Toulouse, Université Paul Sabatier, Toulouse, France.
Electrophoresis. 2010 Jan;31(2):396-402. doi: 10.1002/elps.200900369.
An analytical methodology for quality control analyses of IgG and their impurities is presented using a new UV-LIF (266 nm) detector inside the cassette of a CE instrument and its performance was evaluated. The observed sensitivity was very close to that obtained by silver staining of slab gels (LOD of 25 ng/mL), while the sensitivity of the analysis is 80 times better than with CE/UV absorption (214 nm). Examples of the analysis of pharmaceutical and other commercial IgGs are provided and the kinetics of the reduction of IgG by beta-mercaptoethanol is reported, demonstrating the ease of performing the analysis.
介绍了一种使用 CE 仪器盒内新型 UV-LIF(266nm)检测器进行 IgG 及其杂质质量控制分析的分析方法,并对其性能进行了评估。观察到的灵敏度非常接近平板凝胶银染(LOD 为 25ng/mL)获得的灵敏度,而分析灵敏度比 CE/UV 吸收(214nm)好 80 倍。提供了药物和其他商业 IgG 分析的实例,并报告了β-巯基乙醇还原 IgG 的动力学,证明了进行分析的容易性。