Gibbs N M, Larach D R, Derr J A
Department of Anesthesia, Sir Charles Gairdner Hospital, Nedlands, Western Australia.
Anesthesiology. 1991 Apr;74(4):647-52. doi: 10.1097/00000542-199104000-00004.
The Finapres (FIN) is a new noninvasive blood pressure monitor that provides continuous arterial waveform display with the use of a finger cuff. The authors assessed the accuracy of FIN mean arterial pressure (MAP) measurements relative to simultaneous direct radial arterial pressures in 20 patients undergoing general anesthesia for major elective surgery. Data were collected digitally with the use of RS-232 communications over a total of 16.2 h. The data were processed into 6012 interference-free time samples, each spanning 6 s. The authors determined the difference between FIN and direct MAPs during each time sample. The authors calculated not only the bias of FIN measurements, but also the frequency, magnitude, and duration of discrepancies between simultaneous FIN and direct MAPs. The overall bias of the FIN MAP was -0.5 +/- 1.0 mmHg, which was not significantly different from zero. However, 32.3 +/- 6.2% of all MAP comparisons differed by greater than +/- 10 mmHg, and 5.0 +/- 1.1% differed by greater than +/- 20 mmHg. Moreover, there was an average of one episode every 2 patient-hours when the FIN MAP differed by greater than +/- 20 mmHg for more than 1 min. Although the MAP measured by FIN accurately reflected direct MAPs most of the time, there were occasional discrepancies of different magnitude such that clinical usefulness may be limited in patients in whom continuous accurate blood pressure measurements are essential.
Finapres(FIN)是一种新型无创血压监测仪,通过手指袖带可提供连续动脉波形显示。作者评估了在20例接受重大择期手术全身麻醉的患者中,FIN测量的平均动脉压(MAP)相对于同时直接桡动脉压的准确性。通过RS - 232通信以数字方式总共收集了16.2小时的数据。数据被处理成6012个无干扰时间样本,每个样本跨度为6秒。作者确定了每个时间样本中FIN与直接MAP之间的差异。作者不仅计算了FIN测量的偏差,还计算了同时进行的FIN与直接MAP之间差异的频率、幅度和持续时间。FIN MAP的总体偏差为 -0.5±1.0 mmHg,与零无显著差异。然而,所有MAP比较中有32.3±6.2%的差异大于±10 mmHg,5.0±1.1%的差异大于±20 mmHg。此外,每2个患者小时平均有一次情况,即FIN MAP差异大于±20 mmHg持续超过1分钟。尽管FIN测量的MAP在大多数时候能准确反映直接MAP,但偶尔会出现不同幅度的差异,因此在对连续准确血压测量至关重要的患者中,其临床实用性可能有限。