NHMRC Centre for Clinical Eye Research, Department of Ophthalmology, Flinders University, Flinders Medical Centre, Adelaide, South Australia.
Clin Exp Ophthalmol. 2009 Dec;37(9):836-41; quiz 903-4. doi: 10.1111/j.1442-9071.2009.02181.x.
To determine the sensitivity of manufacturer supported normative algorithms in Heidelberg retinal tomography (HRT) and Stratus optical coherence tomography (OCT) in detecting advanced primary open angle glaucoma.
A total of 157 subjects with advanced primary open angle glaucoma were recruited. The eye with the more severe glaucoma, as judged by mean deviation on Humphrey visual field, was imaged. The total number of optic disc sectors classified as being outside normal limits on Moorfields Regression Analysis (MRA) from the HRT II and HRT 3 software or from the sector and quadrant averages analysis of the Retinal Nerve Fiber Layer OCT scans was compared using the Wilcoxon signed ranks test.
At least one segment in each eye was identified as being outside normal limits by the OCT sector Retinal Nerve Fiber Layer analysis (100% sensitivity). Five eyes were classified as being within normal limits on the OCT quadrant average analysis (96.8% sensitivity). Four eyes were found to be either completely within normal limits or borderline on the HRT II MRA (97.5% sensitivity), of which one was subsequently identified as being outside normal limits by the HRT 3 MRA. The total number of segments classified as being outside normal limits was greater on HRT 3 MRA than HRT II MRA (P < 0.001).
Automated optic nerve imaging devices rarely produce a totally normal result in the presence of advanced primary open angle glaucoma. The Stratus OCT Retinal Nerve Fiber Layer clock hour normative database performs better than other modalities studied. These findings are important in considering the utility of these algorithms for community screening.
确定海德堡视网膜断层扫描仪(HRT)和 Stratus 光学相干断层扫描仪(OCT)中制造商支持的规范算法在检测原发性开角型青光眼进展期时的敏感性。
共招募了 157 例原发性开角型青光眼进展期患者。根据 Humphrey 视野平均偏差判断,对青光眼更严重的眼进行成像。使用 Wilcoxon 符号秩检验比较 HRT II 和 HRT 3 软件的 Moorfields 回归分析(MRA)或视网膜神经纤维层 OCT 扫描的扇形和象限平均值分析中,超出 Moorfields 回归分析(MRA)正常范围的视盘扇形总数。
OCT 扇形视网膜神经纤维层分析(100%敏感性)确定每只眼的至少一个节段超出正常范围。OCT 象限平均值分析(96.8%敏感性)将五只眼归类为正常范围。HRT II MRA(97.5%敏感性)发现四只眼完全或边界正常,其中一只眼随后被 HRT 3 MRA 归类为异常。HRT 3 MRA 分类为异常的节段总数多于 HRT II MRA(P<0.001)。
在原发性开角型青光眼进展期,自动视神经成像设备很少产生完全正常的结果。Stratus OCT 视网膜神经纤维层时钟小时规范数据库的性能优于其他研究的模式。这些发现对于考虑这些算法在社区筛查中的实用性非常重要。