Suppr超能文献

超高交联聚乙烯与传统聚乙烯对氧化锆股骨头的磨损比较:至少 5 年随访。

Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up.

机构信息

Department of Orthopedics Surgery, Kanazawa Medical University, Kahokugun, Japan.

出版信息

J Arthroplasty. 2011 Jan;26(1):45-9. doi: 10.1016/j.arth.2009.11.005. Epub 2010 Jan 22.

Abstract

Highly cross-linked polyethylene (HXLPE) was developed to reduce wear of articular bearing surface in total hip arthroplasty patients. Several studies have shown reduced wear of HXLPE compared with conventional polyethylene; however, these studies had used HXLPE in combination with a Co-Cr metal head. The purpose of this study was to compare the 5-year in vivo wear of HXLPE with that of conventional PE using a zirconia femoral head. Forty-five hips with a Trilogy HXLPE (Zimmer, Warsaw, Ind) were matched and compared with a control group of 20 conventional Trilogy PE hips. The 2-dimensional linear wear rate was significantly less in the HXLPE group between 1 and 5 years postoperation (P < .001). The results show that HXLPE reduces short-term polyethylene wear against not only a Co-Cr head but also a zirconia head.

摘要

超高分子量聚乙烯(HXLPE)的开发旨在减少全髋关节置换患者关节表面的磨损。多项研究表明,与传统聚乙烯相比,HXLPE 的磨损率更低;然而,这些研究均将 HXLPE 与 Co-Cr 金属头联合使用。本研究的目的是使用氧化锆股骨头比较 5 年体内 HXLPE 的磨损与传统 PE 的磨损。45 髋使用 Trilogy HXLPE(Zimmer,印第安纳州华沙),并与对照组 20 髋常规 Trilogy PE 进行匹配和比较。术后 1 至 5 年,HXLPE 组的二维线性磨损率显著降低(P <.001)。结果表明,HXLPE 不仅可以降低 Co-Cr 头,还可以降低氧化锆头的短期聚乙烯磨损。