Hip and Knee Service, Hospital for Special Surgery, New York, New York, USA.
J Arthroplasty. 2011 Jan;26(1):113-9, 119.e1. doi: 10.1016/j.arth.2009.10.018. Epub 2010 Jan 22.
Modularity in sleeved femoral components allows the exchange of the stem without disruption of the fixation between the sleeve and the surrounding bone at revision surgery. Failure to disengage the stem from the sleeve would represent an unnecessary compromise from the intended usefulness of the modular design. We report the results of an examination of 22 modular titanium alloy femoral components retrieved after 0.0 to 8.8 years in vivo. In 7 implants, the stem-sleeve interface could not be disengaged without cutting through the components or using mechanical force. Moderate to severe corrosion was detected in all 7 of these cases. Corrosive surface changes were observed in an additional 6 interfaces. There was no correlation with the length of time that the devices had been implanted. When only the stem is to be revised, orthopedic surgeons should be aware of difficulties in disengagement and anticipate alternative surgical procedures.
套管股骨组件的模块化设计允许在翻修手术中更换股骨柄而不破坏套管与周围骨之间的固定。如果无法将股骨柄从套管中分离出来,将代表对模块化设计的预期用途的不必要妥协。我们报告了对 22 个在体内使用 0.0 至 8.8 年后取出的模块化钛合金股骨组件进行检查的结果。在 7 个植入物中,无法在不切割组件或使用机械力的情况下分离股骨柄-套管界面。在所有这 7 种情况下均检测到中度至重度腐蚀。另外 6 个界面观察到腐蚀性表面变化。这与器械植入的时间长短无关。当仅需修改股骨柄时,矫形外科医生应注意分离的困难,并预计替代手术程序。