Shaiju V S, Sharma S D, Kumar Rajesh, Sarin B
Radiological Physics & Advisory Division, Bhabha Atomic Research Centre, Mumbai - 400 094, India.
J Med Phys. 2009 Jul;34(3):161-6. doi: 10.4103/0971-6203.54851.
Medical cyclotron is a particle accelerator used in producing short lived radiotracers such as (18)F, (11)C, (15)O, (13)N etc. These radiotracers are labeled with suitable pharmaceuticals for use to gather information related to metabolic activity of the cell using Positron Emission Tomography (PET) scan. Target foil rupture is considered one of the major emergency situations during medical cyclotron operations because there is a potential of over exposure to the working personnel. Radiation protection survey of a self-shielded medical cyclotron installation was carried out during normal and emergency conditions. It is found that the induced activity in the target foil increases with its successive usages. As a case study, we have evaluated the emergency handling procedures of GE PETtrace-6 medical cyclotron. Recommendations have also been made to reduce personal exposure while handling the target foil rupture condition such as the use of L-Bench near the target area and participation of experienced personnel.
医用回旋加速器是一种粒子加速器,用于生产如(18)F、(11)C、(15)O、(13)N等短寿命放射性示踪剂。这些放射性示踪剂用合适的药物进行标记,以便利用正电子发射断层扫描(PET)来收集与细胞代谢活性相关的信息。靶箔破裂被认为是医用回旋加速器运行期间的主要紧急情况之一,因为存在工作人员过度暴露的可能性。在正常和紧急情况下对一台自屏蔽式医用回旋加速器装置进行了辐射防护调查。发现靶箔中的感生放射性随着其连续使用而增加。作为一个案例研究,我们评估了GE PETtrace - 6医用回旋加速器的应急处理程序。还提出了在处理靶箔破裂情况时减少个人暴露的建议,如在靶区附近使用L形工作台以及让经验丰富的人员参与。