Proton magnetic resonance spectroscopy of musculoskeletal lesions at 3 T with metabolite quantification.

机构信息

Research Institute and Hospital, National Cancer Center, Goyang-si, Gyeonggi-do 410-769, South Korea.

出版信息

Clin Imaging. 2010 Jan-Feb;34(1):47-52. doi: 10.1016/j.clinimag.2009.03.013.

Abstract

PURPOSE

To evaluate whether proton MR spectroscopy (MRS) at 3 T with metabolite quantification is helpful for characterizing musculoskeletal lesions and to reveal whether the concentration of choline is correlated with the pathologic degree of malignancy.

MATERIAL AND METHODS

Three-tesla MR images and proton MRS data from 27 consecutive patients with surgically proven musculoskeletal lesions were retrospectively analyzed. We analyzed the presence of choline peaks of malignant tumors according to the degree of malignancies, and we compared them with those of benign lesions. The concentrations of choline calculated by means of the linear combination of model spectra were also compared with respect to the degree of malignancy in each group.

RESULTS

The proton MRS had an overall sensitivity of 68.4%, specificity of 87.5%, positive predictive value of 92.3%, and negative predictive value of 53.8% for the detection of choline compounds. The positive detection rate for choline compounds in the primary malignancy group (53.8%) was lower than that of the metastasis group (100%). All false-negative results were shown in the Grade 1 primary malignancy group. There was no difference in the concentration of choline compounds with respect to the pathologic degree of differentiation.

CONCLUSION

MR spectroscopy at 3 T with metabolite quantification is a helpful method that shows high specificity (87.5%) in characterizing musculoskeletal lesions, even though its sensitivity (68.4%) is relatively low. Grade 1 primary malignancies of bone and soft tissue tumor have a high potential for producing false-negative results.

摘要

目的

评估 3T 质子磁共振波谱(MRS)代谢物定量分析是否有助于对肌肉骨骼病变进行特征描述,并探讨胆碱浓度是否与恶性程度的病理分级相关。

材料与方法

回顾性分析 27 例经手术证实的肌肉骨骼病变患者的 3T 磁共振图像和质子 MRS 数据。我们根据恶性程度分析恶性肿瘤的胆碱峰的出现情况,并与良性病变进行比较。通过线性组合模型谱计算的胆碱浓度也与每组的恶性程度进行了比较。

结果

质子 MRS 对胆碱化合物的检测具有总体敏感性 68.4%、特异性 87.5%、阳性预测值 92.3%和阴性预测值 53.8%。原发性恶性肿瘤组(53.8%)的胆碱化合物阳性检出率低于转移组(100%)。所有假阴性结果均显示在 1 级原发性恶性肿瘤组。胆碱化合物的浓度与病理分化程度无关。

结论

3T 磁共振波谱代谢物定量分析是一种有用的方法,在对肌肉骨骼病变进行特征描述方面具有较高的特异性(87.5%),尽管其敏感性(68.4%)相对较低。1 级原发性骨和软组织肿瘤恶性肿瘤具有产生假阴性结果的高风险。