扭转式和混合式超声乳化术后角膜伤口结构与完整性:标准同轴技术与微小切口同轴技术的评估

Corneal wound architecture and integrity after torsional and mixed phacoemulsification: evaluation of standard and microincisional coaxial techniques.

作者信息

Jun Bokkwan, Berdahl John P, Kuo Anthony N, Cummings Thomas J, Kim Terry

机构信息

Erwin Road - Box 3802, Duke University Eye Center, Durham, NC 27710, USA.

出版信息

Ophthalmic Surg Lasers Imaging. 2010 Jan-Feb;41(1):128-34. doi: 10.3928/15428877-20091230-23.

Abstract

BACKGROUND AND OBJECTIVE

To compare the effects of torsional and mixed ultrasound on clear corneal incision architecture, wound integrity, and apposition using standard (2.75 mm) and microincisional (2.2 mm) coaxial phacoemulsification.

SUBJECTS AND METHODS

Twenty human cadaver eyes (4 groups of 5 eyes) underwent simulated coaxial phacoemulsification for 45 seconds of ultrasound time (group 1 = 2.75 mm, 100% torsional; group 2 = 2.2 mm, 70% torsional; group 3 = 2.2 mm, 100% torsional; group 4 = 2.2 mm, mixed ultrasound). All phacoemulsification settings were kept constant across each group. Following phacoemulsification, intraocular pressure (IOP) was cyclically raised and lowered from 0 to 125 mm Hg. Two eyes from each group had India ink placed over the wound and were observed for leakage and for histopathologic examination. Eyes not exposed to India ink (three eyes of each group) were examined using anterior segment optical coherence tomography (OCT) and scanning electron microscopy (SEM).

RESULTS

Wound leakage was evident in one eye from group 1 and no eyes from the other three groups. Histopathologic examination revealed no India ink penetration in any of the eyes studied. Anterior segment OCT showed good wound apposition in each group. SEM demonstrated partially compromised endothelium and Descemet's membrane in all eyes studied from each group.

CONCLUSION

No differences in corneal wound architecture and integrity were observed. Torsional and mixed ultrasound settings do not appear to induce any adverse effects on corneal wound architecture and integrity in standard and microincisional coaxial phacoemulsification techniques.

摘要

背景与目的

使用标准(2.75毫米)和微小切口(2.2毫米)同轴超声乳化技术,比较扭转超声和混合超声对透明角膜切口结构、伤口完整性及对合情况的影响。

对象与方法

20只人尸体眼(4组,每组5只眼)接受模拟同轴超声乳化,超声时间为45秒(第1组 = 2.75毫米,100%扭转;第2组 = 2.2毫米,70%扭转;第3组 = 2.2毫米,100%扭转;第4组 = 2.2毫米,混合超声)。每组的所有超声乳化参数均保持恒定。超声乳化后,眼内压(IOP)在0至125毫米汞柱之间循环升降。每组两只眼在伤口上涂抹印度墨水,观察有无渗漏并进行组织病理学检查。未涂抹印度墨水的眼(每组三只眼)使用眼前节光学相干断层扫描(OCT)和扫描电子显微镜(SEM)进行检查。

结果

第1组有一只眼出现伤口渗漏,其他三组均未出现。组织病理学检查显示,所有研究的眼中均无印度墨水渗透。眼前节OCT显示每组伤口对合良好。SEM显示每组所有研究的眼中内皮和后弹力层均有部分受损。

结论

未观察到角膜伤口结构和完整性存在差异。在标准和微小切口同轴超声乳化技术中,扭转超声和混合超声设置似乎不会对角膜伤口结构和完整性产生任何不良影响。