Primary stability of custom and anatomical uncemented femoral stems: a method for three-dimensional in vitro measurement of implant stability.

作者信息

Østbyhaug Per Olav, Klaksvik Jomar, Romundstad Pål, Aamodt Arild

机构信息

Department of Orthopedic Surgery, Trondheim University Hospital, 7006 Trondheim, Norway.

出版信息

Clin Biomech (Bristol). 2010 May;25(4):318-24. doi: 10.1016/j.clinbiomech.2009.12.012. Epub 2010 Feb 2.

Abstract

BACKGROUND

Lack of primary stability of cementless hip stems prevents bone ingrowth and may lead to loosening of the stem. Direct measures of the implant stability require drilled holes in the bone at the measuring site. These holes weaken the cortical bone, limit the number of possible measuring points and inhibit other biomechanical measurements. This in vitro study aimed to develop a method for indirect measurement of primary stability of femoral stems, leaving the specimen intact. The method was used to compare the primary stability of two uncemented femoral stems with different proximal fit and fill and different stem length.

METHODS

An in vitro method for indirect full three-dimensional measurement of implant-bone interface motion was developed. Uncemented customized (n=10) and anatomical stems (n=10) were inserted in human cadaver femora and the primary stability during one leg stance and stair climbing was measured.

FINDINGS

The method had high precision, and the errors due to necessary assumption of rigid body components were minimal. The customized stem with optimal proximal fit and fill provided the best initial stability for rotation in retroversion. The anatomical stem with longer stem length was more resistant to permanent rotation in varus.

INTERPRETATION

During stem design development the primary stability can be measured at all wanted measuring sites with the presented method, leaving the specimen intact for further analyses.

摘要

背景

非骨水泥型髋关节柄缺乏初始稳定性会阻碍骨长入,并可能导致柄松动。直接测量植入物稳定性需要在测量部位的骨上钻孔。这些孔会削弱皮质骨,限制可能的测量点数量,并妨碍其他生物力学测量。这项体外研究旨在开发一种间接测量股骨干初始稳定性的方法,使标本保持完整。该方法用于比较两种具有不同近端贴合度和填充度以及不同柄长度的非骨水泥型股骨干的初始稳定性。

方法

开发了一种用于间接全三维测量植入物-骨界面运动的体外方法。将定制的非骨水泥型(n = 10)和解剖型柄(n = 10)插入人体尸体股骨中,并测量单腿站立和爬楼梯期间的初始稳定性。

结果

该方法具有很高的精度,由于刚体组件的必要假设导致的误差最小。具有最佳近端贴合度和填充度的定制柄为后倾旋转提供了最佳的初始稳定性。柄长度较长的解剖型柄在内翻时更能抵抗永久性旋转。

解读

在柄设计开发过程中,可以使用本文提出的方法在所有需要的测量部位测量初始稳定性,使标本保持完整以便进一步分析。