Rapid determination of glucose-6-phosphate dehydrogenase (G6PD) status is desirable when it is necessary to use a drug contraindicated in G6PD-deficient persons, such as use of primaquine for malaria prevention or treatment. The purpose of this study was to compare a new, rapid, qualitative enzyme chromatographic test for deficiency of G6PD to a standard reference method. Samples from 196 G6PD-normal persons and 50 G6PD-deficient persons were evaluated. The sensitivity of the experimental rapid test was 0.98 and the specificity was 0.98 using specimens preserved in heparin, and 0.98 and 0.97, respectively, for specimens preserved in EDTA. Positive and negative predictive values were 0.72 and 1.00, respectively, for the test for heparinized specimens and 0.65 and 1.00, respectively, for the EDTA-preserved samples. This rapid test for G6PD deficiency is a sensitive method for screening of G6PD deficiency that requires minimal training and equipment and enables rapid identification of G6PD-deficient persons.
当需要使用 G6PD 缺乏症患者禁用的药物时,如使用伯氨喹预防或治疗疟疾,快速确定葡萄糖-6-磷酸脱氢酶(G6PD)的状态是理想的。本研究的目的是比较一种新的快速定性酶色谱试验与标准参考方法检测 G6PD 缺乏症的能力。评估了 196 名 G6PD 正常人和 50 名 G6PD 缺乏症患者的样本。使用肝素保存的标本时,实验性快速试验的灵敏度为 0.98,特异性为 0.98,EDTA 保存的标本的灵敏度和特异性分别为 0.98 和 0.97。肝素化标本试验的阳性和阴性预测值分别为 0.72 和 1.00,EDTA 保存标本的阳性和阴性预测值分别为 0.65 和 1.00。这种快速检测 G6PD 缺乏症的方法是一种用于 G6PD 缺乏症筛查的敏感方法,它需要最少的培训和设备,并能够快速识别 G6PD 缺乏症患者。