Does a flat panel detector reduce the patient radiation dose in interventional cardiology?

作者信息

Dragusin O, Breisch R, Bokou C, Beissel J

机构信息

Entente des Hôpitaux Luxembourgeois, 5 rue des Mérovingiens, L-8070 Bertrange, Luxembourg.

出版信息

Radiat Prot Dosimetry. 2010 Apr-May;139(1-3):266-70. doi: 10.1093/rpd/ncq008. Epub 2010 Feb 5.

Abstract

An internal audit has been performed to evaluate the patient radiation dose in the clinical use of X-ray cardiac flat panel detector (FPD) systems. Fluoroscopy and image acquisition (cine mode) programs are optimised for low-dose settings. The diagnostic and therapeutic interventional cardiac procedures are performed at 6 pulse s(-1) for fluoroscopy and 15 frames s(-1) for image acquisition. The FPD entrance dose is configured for 29 nGy pulse(-1) in the fluoroscopy mode and 0.10 microGy frame(-1) in the cine mode. From the data collected, the following local reference levels for coronary angiography and angioplasty have been obtained: kerma-area products 23 and 44 Gy cm(2), cumulative doses 376 and 776 mGy, fluoroscopy times 5 and 15.5 min and cine images 617 and 1163 images.

摘要

已进行内部审计,以评估X射线心脏平板探测器(FPD)系统临床应用中的患者辐射剂量。透视和图像采集(电影模式)程序针对低剂量设置进行了优化。诊断性和治疗性心脏介入手术在透视时以6脉冲/秒、图像采集时以15帧/秒进行。FPD入口剂量在透视模式下配置为29 nGy/脉冲,在电影模式下配置为0.10 μGy/帧。根据收集的数据,已获得以下冠状动脉造影和血管成形术的本地参考水平:比释动能面积乘积分别为23和44 Gy cm²,累积剂量分别为376和776 mGy,透视时间分别为5和15.5分钟,电影图像分别为617和1163幅。