髋臼杯边缘设计对陶瓷-陶瓷髋关节假体边缘加载时接触应力的影响。

Influence of acetabular cup rim design on the contact stress during edge loading in ceramic-on-ceramic hip prostheses.

机构信息

School of Mechanical Engineering, The University of Leeds, Leeds, United Kingdom.

出版信息

J Arthroplasty. 2011 Jan;26(1):131-6. doi: 10.1016/j.arth.2009.10.019. Epub 2010 Feb 9.

Abstract

The purpose of the study was to investigate the contact stresses in 3 different acetabular cup rim designs (new, worn, chamfer) during edge loading, after microseparation of ceramic on ceramic hip prostheses. A 3-dimensional finite element analysis was conducted for a 28-mm diameter alumina ceramic bearing with a radial clearance of 40 μm using a normal load of 2500 N under edge loading. At a separation distance of 250 μm, the maximum tensile stress in the "new" design was of similar magnitude to the flexural strength of the alumina material that supports the localized breakdown (stripe wear) of the acetabular cup surface observed clinically. Introducing a 2.5-mm radius chamfer should reduce the maximum tensile stress in the region of 60%.

摘要

本研究的目的是研究在陶瓷-陶瓷髋关节假体发生微分离后,边缘加载时 3 种不同髋臼杯边缘设计(新的、磨损的、倒角的)的接触应力。通过对直径为 28mm、具有 40μm 径向间隙的氧化铝陶瓷轴承进行三维有限元分析,在边缘加载下使用 2500N 的法向载荷,分离距离为 250μm。在“新”设计中,最大拉伸应力与支撑髋臼杯表面局部失效(条纹磨损)的氧化铝材料的抗弯强度相当,临床上观察到了这种失效。引入 2.5mm 半径的倒角可将该区域的最大拉伸应力降低 60%左右。