Thomas P R, Butler C M, Bowman J, Grieve N W, Bennett C E, Taylor R S, Thomas M H
Julie Andrews Vascular Unit, St. Peter's Hospital, Chertsey, Surrey, UK.
Br J Surg. 1991 Feb;78(2):207-9. doi: 10.1002/bjs.1800780225.
Light reflection rheography is a simple non-invasive technique for assessing venous function in the leg. One hundred and twenty-four patients referred for venography with a clinically suspected deep venous thrombosis were investigated by light reflection rheography to determine the accuracy of the technique in diagnosing acute thrombosis. In half of the patients venography confirmed a deep venous thrombosis. Light reflection rheography had a sensitivity of 92 per cent and a specificity of 84 per cent in detecting acute thrombosis. The technique had a negative predictive value of 92 per cent in selecting those patients with no thrombosis. Light reflection rheography can be performed at the bedside or in the radiography department which makes it a suitable technique for screening patients with suspected deep venous thrombosis.
光反射血流描记术是一种用于评估腿部静脉功能的简单无创技术。对124例因临床怀疑深静脉血栓形成而转诊进行静脉造影的患者,采用光反射血流描记术进行研究,以确定该技术在诊断急性血栓形成中的准确性。在一半的患者中,静脉造影证实存在深静脉血栓形成。光反射血流描记术在检测急性血栓形成方面的敏感性为92%,特异性为84%。该技术在筛选无血栓形成的患者时,阴性预测值为92%。光反射血流描记术可在床边或放射科进行,这使其成为筛查疑似深静脉血栓形成患者的合适技术。