Development, validation, and certification by isotope dilution gas chromatography-mass spectrometry of lyophilized human serum reference materials for cortisol (CRM 192 and 193) and progesterone (CRM 347 and 348).

作者信息

Thienpont L, Siekmann L, Lawson A, Colinet E, De Leenheer A

机构信息

Laboratoria voor Medische Biochemie en voor Klinische Analyse, RUG, Ghent, Belgium.

出版信息

Clin Chem. 1991 Apr;37(4):540-6.

PMID:2015668
Abstract

The Community Bureau of Reference of the European Communities has produced four batches of lyophilized serum Certified Reference Materials, two for cortisol (CRM 192 and 193) and two for progesterone (CRM 347 and 348). For cortisol, one of the pools consisted of serum from healthy blood donors, whereas the second batch was supplemented with pure cortisol. The progesterone Reference Materials contained only endogenous hormone concentrations. Assessment of vial-to-vial variability in the cortisol and progesterone concentrations showed no between-sample inhomogeneity, and the materials were stable. The quality of the materials was therefore considered sufficient for certification of the values for the cortisol and progesterone concentrations by a collaborative study involving several laboratories from the European Communities, using isotope dilution gas chromatography-mass spectrometry. Inaccuracy in reconstitution of the lyophilized materials was less than 0.3%; imprecision of sampling was less than 0.2%. For determinations of cortisol and progesterone concentrations, the mean within-laboratory coefficients of variation (CVs) were 1.76% (CRM 192), 1.19% (CRM 193), 1.64% (CRM 347), and 1.75% (CRM 348). The between-laboratory CVs were greater: CRM 192, 1.79%; CRM 193, 1.48%; CRM 347, 2.08%; and CRM 348, 2.16%. The concentrations in the reconstituted Reference Materials were certified to be 273 nmol/L in CRM 192 and 763 nmol/L in CRM 193 for cortisol and 10.13 nmol/L in CRM 347 and 40.3 nmol/L in CRM 348 for progesterone. Uncertainties at the 0.95 confidence level--6 (CRM 192), 14 (CRM 193), 0.21 (CRM 347), and 1.0 nmol/L (CRM 348)--were considered compatible with the intended use of the materials.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验