BACKGROUND

Current methods for detecting atrial fibrillation (AF) have limited diagnostic yield. Continuous monitoring with automatic arrhythmia detection and classification may improve detection of symptomatic and asymptomatic AF and subsequent patient treatment. The study purpose was to quantify the performance of the first implantable leadless cardiac monitor (ICM) with dedicated AF detection capabilities.

METHODS AND RESULTS

Patients (n=247) with an implanted ICM (Reveal XT, Medtronic Inc, Minneapolis, Minn) who were likely to present with paroxysmal AF were selected. A special Holter device stored 46 hours of subcutaneously recorded ECG, ICM markers, and 2 surface ECG leads. The ICM automatic arrhythmia classification was compared with the core laboratory classification of the surface ECG. Of the 206 analyzable Holter recordings collected, 76 (37%) contained at least 1 episode of core laboratory classified AF. The sensitivity, specificity, positive predictive value, and negative predictive value for identifying patients with any AF were 96.1%, 85.4%, 79.3%, and 97.4%, respectively. The AF burden measured with the ICM was very well correlated with the reference value derived from the Holter (Pearson coefficient=0.97). The overall accuracy of the ICM for detecting AF was 98.5%.

CONCLUSIONS

In this ICM validation study, the dedicated AF detection algorithm reliably detected the presence or absence of AF and the AF burden was accurately quantified. The ICM is a promising new diagnostic and monitoring tool for the clinician to treat AF patients independently of symptoms. Long-term studies are needed to evaluate the clinical benefits of the technology.

摘要

背景

目前用于检测心房颤动(AF)的方法诊断率有限。通过自动心律失常检测和分类的连续监测可能会提高有症状和无症状 AF 的检出率,以及随后的患者治疗效果。本研究的目的是量化具有专用 AF 检测功能的首款植入式无导线心脏监测器(ICM)的性能。

方法和结果

选择了植入 ICM(Medtronic Inc,明尼苏达州明尼阿波利斯市的 Reveal XT)且可能出现阵发性 AF 的患者(n=247)。特殊的动态心电图仪储存了 46 小时的皮下记录心电图、ICM 标记和 2 个体表心电图导联。ICM 自动心律失常分类与体表心电图的核心实验室分类进行了比较。在收集的 206 个可分析的动态心电图记录中,76 个(37%)至少包含 1 次核心实验室分类的 AF 发作。识别任何 AF 患者的敏感性、特异性、阳性预测值和阴性预测值分别为 96.1%、85.4%、79.3%和 97.4%。ICM 测量的 AF 负荷与动态心电图仪参考值高度相关(Pearson 系数=0.97)。ICM 检测 AF 的总体准确性为 98.5%。

结论

在这项 ICM 验证研究中,专用的 AF 检测算法可靠地检测了 AF 的存在或不存在,并且准确地量化了 AF 负荷。ICM 是一种有前途的新诊断和监测工具,可供临床医生独立于症状来治疗 AF 患者。需要进行长期研究来评估该技术的临床获益。