Absolute reference calorimeter for measuring high power laser pulses.

作者信息

Franzen D L, Schmidt L B

出版信息

Appl Opt. 1976 Dec 1;15(12):3115-22. doi: 10.1364/AO.15.003115.

Abstract

A calorimeter for making absolute energy measurements of high power laser pulses is described. The calorimeter, based on volume absorption in a solid, is calibrated electrically and requires no window or vacuum environment. An error analysis is included giving the systematic and random errors of the instrument for a laser measurement. Briefly, the following performance is typical of the 32-mm x 32-mm aperture calorimeter: range 0.4-15-J; random error +/-0.2% (one standard deviation); systematic error +/-2.3%; and an upper operational limit of 3 J/cm(2). Most of the volume absorber documentation is applicable for 1.06 microm; however, the calorimeter should be useful from the near ir through the visible. Absorbers for use with CO(2) lasers in the 9-11-microm range are also discussed.