Fialon P, Macaigne F, Becker M, Boisseau M R, Cazenave J, Ripert C
Laboratoire de Biologie, Hôpital Saint-André, Bordeaux, France.
Pathol Biol (Paris). 1991 Feb;39(2):122-5.
Blood cell counts were performed on blood samples from 37 patients with imported malaria using three different blood analyzers (Coulter STKR, Coulter VCS and Technicon H1). Results were controlled by direct microscopic examination. Anemia, leukopenia, thrombocytopenia, or abnormalities of the leukocyte differential count were found in 32, 24, 30 and 92% of patients, respectively. The automatic analyzers gave alert messages for 70 to 75% of specimens, including specimens from ten patients with low parasitemias. These abnormalities should prompt careful analysis of blood smears when drug-resistance is suspected.
使用三种不同的血液分析仪(库尔特STKR、库尔特VCS和Technicon H1)对37例输入性疟疾患者的血样进行血细胞计数。结果通过直接显微镜检查进行对照。分别在32%、24%、30%和92%的患者中发现贫血、白细胞减少、血小板减少或白细胞分类计数异常。自动分析仪对70%至75%的标本发出警报信息,包括来自10例低疟原虫血症患者的标本。当怀疑耐药时,这些异常情况应促使对血涂片进行仔细分析。