Division of Imaging and Applied Mathematics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug, 10903 New Hampshire Avenue, WO 62-3116, Silver Spring, Maryland 20993, USA.
Med Phys. 2010 Jan;37(1):74-81. doi: 10.1118/1.3265879.
The necessity for standard instrumentation and measurements of color that are repeatable and reproducible is the major motivation behind this work. Currently, different instrumentation and methods can yield very different results when measuring the same feature such as color uniformity or color difference. As color increasingly comes into play in medical imaging diagnostics, display color will have to be quantified in order to assess whether the display should be used for imaging purposes. The authors report on the characterization of three novel probes for measuring display color with minimal contamination from screen areas outside the measurement spot or from off-normal emissions. They compare three probe designs: A modified small-spot luminance probe and two conic probe designs based on black frusta.
To compare the three color probe designs, spectral and luminance measurements were taken with specialized instrumentation to determine the luminance changes and color separation abilities of the probes. The probes were characterized with a scanning slit method, veiling glare, and a moving laser and LED arrangement. The scanning slit measurement was done using a black slit plate over a white line on an LCD monitor. The luminance was measured in 1 mm increments from the center of the slit to +/- 15 mm above and below the slit at different distances between the probe and the slit. The veiling glare setup consisted of measurements of the luminance of a black spot pattern with a white disk of radius of 100 mm as the black spot increases in 1 mm radius increments. The moving LED and laser method consisted of a red and green light orthogonal to the probe tip for the light to directly shine into the probe. The green light source was moved away from the red source in 1 cm increments to measure color stray-light contamination at different probe distances.
The results of the color testing using the LED and laser methods suggest a better performance of one of the frusta probes at shorter distances between the light sources, which translates to less contamination. The tails of the scans indicate the magnitude of the spread in signal due to light from areas outside the intended measurement spot. The measurements indicate a corresponding glare factor for a large spot of 140, 500, and 2000 for probe A, B1, and B2, respectively. The dual-laser setup suggests that color purity can be maintained up to a few tens of millimeters outside the measurement spot.
The comparison shows that there are significant differences in the performance of each probe design, and that those differences have an effect on the measured quantity used to quantify display color. Different probe designs show different measurements of the level of light contamination that affects the quantitative color determination.
需要对颜色进行可重复和可再现的标准化仪器和测量,这是这项工作的主要动机。目前,当测量相同的特征(如颜色均匀性或颜色差异)时,不同的仪器和方法可能会产生非常不同的结果。随着颜色在医学成像诊断中的作用越来越大,为了评估显示器是否可用于成像目的,必须对显示颜色进行量化。作者报告了三种新型探头的特性,这些探头可用于测量显示器颜色,同时最大限度地减少来自测量点以外的屏幕区域或来自非正常发射的污染。他们比较了三种探头设计:一种改进的小光斑亮度探头和两种基于黑色截头圆锥的圆锥探头设计。
为了比较三种颜色探头设计,使用专用仪器进行了光谱和亮度测量,以确定探头的亮度变化和颜色分离能力。使用扫描狭缝法、眩光和移动激光和 LED 装置对探头进行了特性描述。扫描狭缝测量是在液晶显示器上的白线和黑线之间的白色狭缝板上进行的。亮度是在探针与狭缝之间不同距离处,从狭缝中心以 1 毫米的增量测量到狭缝上方和下方 15 毫米处。眩光设置包括用直径为 100 毫米的白色圆盘测量黑色斑点图案的亮度,随着黑色斑点以 1 毫米半径的增量增加。移动 LED 和激光方法包括一个红色和绿色光源,与探头尖端正交,以便光直接照射到探头中。红色光源以 1 厘米的增量远离绿色光源,以测量不同探头距离处的颜色杂散光污染。
使用 LED 和激光方法进行的颜色测试结果表明,在光源之间的距离较短时,一种截头圆锥探头的性能更好,这意味着污染较少。扫描的尾部指示由于测量点以外区域的光导致信号扩散的幅度。测量结果表明,对于大斑点,探针 A、B1 和 B2 的相应眩光因子分别为 140、500 和 2000。双激光设置表明,在测量点外几十毫米的范围内可以保持颜色纯度。
比较表明,每个探头设计的性能存在显著差异,这些差异会影响用于量化显示颜色的测量量。不同的探头设计对影响定量颜色确定的光污染水平有不同的测量。