Unit of Medical Physics, Azienda Ospedaliero-Universitaria Pisana, 56126 Pisa, Italy.
Med Phys. 2010 Jan;37(1):303-10. doi: 10.1118/1.3271130.
Although in EPI-fMRI analyses typical acquisition parameters (TR, TE, matrix, slice thickness, etc.) are generally employed, various readout bandwidth (BW) values are used as a function of gradients characteristics of the MR scanner. Echo spacing (ES) is another fundamental parameter of EPI-fMRI acquisition sequences but the employed ES value is not usually reported in fMRI studies. In the present work, the authors investigated the effect of ES and BW on basic performances of EPI-fMRI sequences in terms of temporal stability and overall image quality of time series acquisition.
EPI-fMRI acquisitions of the same water phantom were performed using two clinical MR scanner systems (scanners A and B) with different gradient characteristics and functional designs of radiofrequency coils. For both scanners, the employed ES values ranged from 0.75 to 1.33 ms. The used BW values ranged from 125.0 to 250.0 kHz/64pixels and from 78.1 to 185.2 kHz/64pixels for scanners A and B, respectively. The temporal stability of EPI-fMRI sequence was assessed measuring the signal-to-fluctuation noise ratio (SFNR) and signal drift (DR), while the overall image quality was assessed evaluating the signal-to-noise ratio (SNR(ts)) and nonuniformity (NU(ts)) of the time series acquisition.
For both scanners, no significant effect of ES and BW on signal drift was revealed. The SFNR, NU(ts) and SNR(ts) values of scanner A did not significantly vary with ES. On the other hand, the SFNR, NU(ts), and SNR(ts) values of scanner B significantly varied with ES. SFNR (5.8%) and SNR(ts) (5.9%) increased with increasing ES. SFNR (25% scanner A, 32% scanner B) and SNR(ts) (26.2% scanner A, 30.1% scanner B) values of both scanners significantly decreased with increasing BW. NU(ts) values of scanners A and B were less than 3% for all BW and ES values. Nonetheless, scanner A was characterized by a significant upward trend (3% percentage of variation) of time series nonuniformity with increasing BW while NU(ts) of scanner B significantly increased (19% percentage of variation) with increasing ES.
Temporal stability (SFNR and DR) and overall image quality (NU(ts) and SNR(ts)) of EPI-fMRI time series can significantly vary with echo spacing and readout bandwidth. The specific pattern of variation may depend on the performance of each single MR scanner system in terms of gradients characteristics, EPI sequence calibrations (eddy currents, shimming, etc.), and functional design of radiofrequency coil. Our results indicate that the employment of low BW improves not only the signal-to-noise ratio of EPI-fMRI time series but also the temporal stability of functional acquisitions. The use of minimum ES values is not entirely advantageous when the MR scanner system is characterized by gradients with low performances and suboptimal EPI sequence calibration. Since differences in basic performances of MR scanner system are potential source of variability for fMRI activation, phantom measurements of SFNR, DR, NU(ts), and SNR(ts) can be executed before subjects acquisitions to monitor the stability of MR scanner performances in clinical group comparison and longitudinal studies.
尽管在 EPI-fMRI 分析中通常采用典型的采集参数(TR、TE、矩阵、切片厚度等),但不同的读出带宽(BW)值是作为磁共振扫描仪梯度特性的函数使用的。回波间隔(ES)是 EPI-fMRI 采集序列的另一个基本参数,但 fMRI 研究中通常不报告所采用的 ES 值。在本工作中,作者研究了 ES 和 BW 对 EPI-fMRI 序列基本性能的影响,包括时间稳定性和时间序列采集的整体图像质量。
使用具有不同梯度特性和射频线圈功能设计的两台临床磁共振扫描仪系统(扫描仪 A 和 B)对相同的水模体进行 EPI-fMRI 采集。对于两台扫描仪,所采用的 ES 值范围为 0.75 至 1.33ms。所采用的 BW 值范围分别为扫描仪 A 的 125.0 至 250.0kHz/64 像素和扫描仪 B 的 78.1 至 185.2kHz/64 像素。通过测量信号波动噪声比(SFNR)和信号漂移(DR)来评估 EPI-fMRI 序列的时间稳定性,通过评估时间序列采集的信号噪声比(SNR(ts))和非均匀性(NU(ts))来评估整体图像质量。
对于两台扫描仪,ES 和 BW 对信号漂移均无显著影响。扫描仪 A 的 SFNR、NU(ts)和 SNR(ts)值不随 ES 显著变化。另一方面,扫描仪 B 的 SFNR、NU(ts)和 SNR(ts)值随 ES 显著变化。SFNR(5.8%)和 SNR(ts)(5.9%)随 ES 增加而增加。SFNR(扫描仪 A 为 25%,扫描仪 B 为 32%)和 SNR(ts)(扫描仪 A 为 26.2%,扫描仪 B 为 30.1%)值随 BW 增加而显著降低。扫描仪 A 和 B 的 NU(ts)值均小于 3%,对于所有 BW 和 ES 值。尽管如此,扫描仪 A 的时间序列非均匀性随 BW 呈显著上升趋势(变化 3%),而扫描仪 B 的 NU(ts)随 ES 显著增加(变化 19%)。
EPI-fMRI 时间序列的时间稳定性(SFNR 和 DR)和整体图像质量(NU(ts)和 SNR(ts))可能随回波间隔和读出带宽显著变化。特定的变化模式可能取决于每台磁共振扫描仪系统在梯度特性、EPI 序列校准(涡流、调谐等)和射频线圈功能设计方面的性能。我们的结果表明,采用低 BW 不仅可以提高 EPI-fMRI 时间序列的信噪比,还可以提高功能采集的时间稳定性。当磁共振扫描仪系统具有性能较低和 EPI 序列校准不佳的梯度时,采用最小 ES 值并不完全有利。由于磁共振扫描仪系统基本性能的差异是 fMRI 激活变异性的潜在来源,因此可以在对受试者进行采集之前在体模上执行 SFNR、DR、NU(ts)和 SNR(ts)的测量,以监测磁共振扫描仪在临床组比较和纵向研究中的性能稳定性。