Vitreoretinal traction created by conventional cutters during vitrectomy.

机构信息

Eye Concepts, Doheny Eye Institute, Los Angeles, California 90033-4500, USA.

出版信息

Ophthalmology. 2010 Jul;117(7):1387-92.e2. doi: 10.1016/j.ophtha.2009.11.004. Epub 2010 Feb 21.

Abstract

PURPOSE

To quantify the vitreous traction created by pneumatic (20-, 23-, and 25-gauge) and electric cutters (20- and 25-gauge) during vitrectomy.

DESIGN

Experimental study.

PARTICIPANTS

This is a preclinical study, and no patients were involved.

METHODS

Five separate cutters for each drive mechanism and gauge (20-, 23-, and 25-gauge pneumatic and 20- and 25-gauge electric) were used for each cut speed at predetermined aspiration rates. The retinal layers of fresh porcine eyes were transfixed with a 0.15-mm steel wire and fixed to the load cell of a strain gauge. The cutter to be assessed was introduced into the eye by a micromanipulator at a 45-degree angle adjacent to the retina. The traction force was determined and evaluated at a distance of 3 and 5 mm from the retina with different vacuum and cut rates.

MAIN OUTCOME MEASURES

Vacuum, cut speed, force (dynes), distance (millimeters), vitreoretinal traction, and gauge.

RESULTS

The 20-, 23-, and 25-gauge pneumatic cutters have a range of traction from 2.06 to 37.22 dynes, 3.85 to 15.38 dynes, 5.13 to 27.91 dynes, respectively. The 20- and 25-gauge electric cutters have a range of traction from 3.60 to 41.78 dynes and 5.28 to 27.91 dynes, respectively. All results are related to distance, cut, and aspiration rate. With an increase of 100 mmHg of vacuum/aspiration, the traction increased from 7.89 to 3.14 dynes (e.g., 4.96 for 20-gauge pneumatic). The traction decreased as the cut rate was increased, from 5.71 to 2.51 dynes (e.g., 3.41 for 20-gauge pneumatic).

CONCLUSIONS

The results indicate that retinal traction increased with increasing aspiration vacuum and proximity to the retina; conversely, retinal traction decreased with increasing cut rate. The present study demonstrates that the effects of aspiration, distance from the retina, and cut rate are crucial factors in the amount of retinal traction created by vitreous cutters.

FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references.

摘要

目的

量化玻璃体切割术中气动(20、23 和 25 号)和电动切割器(20 和 25 号)产生的玻璃体牵引力。

设计

实验研究。

参与者

这是一项临床前研究,没有涉及患者。

方法

为每个切割速度和每个驱动机制和规格(20、23 和 25 号气动和 20 和 25 号电动)使用了五个单独的切割器,以预定的抽吸率进行。新鲜猪眼的视网膜层用 0.15 毫米的钢丝固定,并固定在应变计的称重传感器上。评估用的切割器通过微操纵器以 45 度角引入眼内,与视网膜相邻。在距离视网膜 3 和 5 毫米处,用不同的真空和切割速度确定和评估牵引力。

主要观察指标

真空度、切割速度、力(达因)、距离(毫米)、玻璃体视网膜牵引和规格。

结果

20、23 和 25 号气动切割器的牵引力范围分别为 2.06 至 37.22 达因、3.85 至 15.38 达因和 5.13 至 27.91 达因。20 和 25 号电动切割器的牵引力范围分别为 3.60 至 41.78 达因和 5.28 至 27.91 达因。所有结果均与距离、切割和抽吸率有关。真空/抽吸增加 100mmHg 时,牵引力从 7.89 增加到 3.14 达因(例如,20 号气动的 4.96)。随着切割速度的增加,牵引力从 5.71 减少到 2.51 达因(例如,20 号气动的 3.41)。

结论

结果表明,视网膜牵引力随抽吸真空度和接近视网膜的增加而增加;相反,随着切割速度的增加,视网膜牵引力减小。本研究表明,抽吸、距视网膜的距离和切割速度的影响是玻璃体切割器产生的视网膜牵引力的关键因素。