Suppr超能文献

一种用于量化玻璃体切割术中牵引力的新方法。

Novel method to quantify traction in a vitrectomy procedure.

机构信息

Eye Concepts Laboratory, Los Angeles, CA, USA.

出版信息

Br J Ophthalmol. 2010 Sep;94(9):1226-9. doi: 10.1136/bjo.2009.166637. Epub 2010 Jun 10.

Abstract

AIM

To report a novel method to quantify traction applied to the retina using vitreous cutters during pars plana vitrectomy.

METHODS

Fresh porcine eyes were positioned in a specially developed holder and transfixed to the retinal layers with a wire and the other end fixed to the load cell of a strain gauge. Five separate 20-gauge electrical drive mechanism vitrectors were introduced into the eye at a 45 degrees angle and positioned at a distance of either 3 or 5 mm from the retina. Data from the strain gauge were acquired and the traction force computed.

RESULTS

The analysis revealed that the vitreoretinal traction increased by 7.90 dynes for each 100 mm Hg increase in vacuum (p<0.05). The traction forces decreased by 2.51 dynes for each 500 cuts per minute increased (p<0.05) and the traction force increased by 2.17 dynes at 3 mm compared with 5 mm (p<0.05).

CONCLUSION

The traction was directly proportional to the aspiration vacuum and inversely proportional to the cut rate. The cutter traction force increased with proximity to the retina.

摘要

目的

报告一种在进行标准经睫状体平坦部玻璃体切割术时通过玻璃体切割器定量施加于视网膜牵引力的新方法。

方法

新鲜猪眼被放置在特制的固定器中,通过金属丝将视网膜的各层固定,并将另一端固定于应变片的称重传感器上。将 5 个单独的 20 号电驱动玻璃体切割器以 45 度角引入眼内,距离视网膜 3 或 5mm 处定位。采集应变片的数据并计算牵引力。

结果

分析表明,真空度每增加 100mmHg,玻璃体视网膜牵引力增加 7.90 达因(p<0.05)。每分钟切割次数每增加 500 次,牵引力减少 2.51 达因(p<0.05),与 5mm 相比,3mm 处的牵引力增加 2.17 达因(p<0.05)。

结论

牵引力与吸引真空度成正比,与切割速度成反比。切割器的牵引力随与视网膜的接近程度而增加。