Dipartimento di Bioingegneria, Politecnico di Milano University, Italy.
Physiol Meas. 2010 Apr;31(4):N11-24. doi: 10.1088/0967-3334/31/4/N01. Epub 2010 Feb 24.
Spirometry is the gold standard to determine the presence of airflow obstruction, but it requires volitional participation and needs qualified supervision. The forced oscillation technique (FOT) measures respiratory input impedance (Zrs) during spontaneous breathing and it could be useful for unsupervised monitoring of airway obstruction. We developed a FOT device for home monitoring of Zrs which transmits the data through the Internet. Its accuracy, stability and reliability were evaluated in a pilot study measuring the Zrs in the unsupervised self-measurements of five healthy subjects. Finally, to explore the applicability of the concept, 36 consecutive daily home measurements were recorded from one healthy subject and one chronic obstructive pulmonary disease (COPD) patient. The accuracy of the device fulfilled FOT guidelines, and the reliability test showed a mean discrepancy of resistance of 0.10 +/- 0.01 cmH(2)O s L(-1). The data from the healthy subjects demonstrated high repeatability in assessing Zrs. The measurements on the healthy subjects and the patient with COPD suggest the feasibility of unsupervised FOT measurements. The healthy subjects showed minimal daily variations in Zrs, whereas the patient with COPD had large differences in mean values and important fluctuations over day-to-day measurements. The results of the pilot study demonstrate that unsupervised home monitoring of Zrs using the FOT yields accurate and reproducible data. It could provide new insights into the dynamics of airway obstruction and improve the understanding and management of obstructive diseases.
肺量测定法是确定气流阻塞存在的金标准,但它需要自愿参与并且需要合格的监督。强迫振荡技术(FOT)在自主呼吸期间测量呼吸输入阻抗(Zrs),并且它可能对气道阻塞的非监督监测有用。我们开发了一种用于家庭监测 Zrs 的 FOT 设备,该设备通过互联网传输数据。在一项针对五名健康受试者的非监督自我测量中,对其准确性、稳定性和可靠性进行了评估。最后,为了探索该概念的适用性,对一名健康受试者和一名慢性阻塞性肺疾病(COPD)患者进行了 36 次连续的日常家庭测量。该设备的准确性符合 FOT 指南,可靠性测试显示阻力的平均差异为 0.10 +/- 0.01 cmH(2)O s L(-1)。健康受试者评估 Zrs 的重复性很高。对健康受试者和 COPD 患者的测量表明,非监督 FOT 测量是可行的。健康受试者的 Zrs 每日变化最小,而 COPD 患者的平均数值差异较大,并且每日测量之间波动较大。该初步研究的结果表明,使用 FOT 进行非监督家庭监测 Zrs 可提供准确且可重复的数据。它可以提供对气道阻塞动态的新见解,并改善对阻塞性疾病的理解和管理。