To compare intra-individual contrast enhancement in multi-detector-row computed tomography (MDCT) using contrast media (CM) containing 300, 370 and 400 mg iodine per ml (mgI/ml).
Six pigs underwent repeated chest MDCT using three different CM (iopromide 300, iopromide 370, iomeprol 400). An identical iodine delivery (IDR) rate of 1.5 gI/s and a constant total iodine dose of 300 mg/kg body weight were used. Dynamic CT were acquired at the level of the pulmonary artery, and the ascending and descending aorta. After the time enhancement curves were computed, the pulmonary and aortic peak enhancement, time to peak and plateau time above 300 HU were calculated.
Intra-individual peak contrast enhancement was significantly higher for the 300 mgI/ml contrast medium compared with the 370 and 400 mgI/ml media: pulmonary trunk 595 HU vs 516 HU (p = 0.0093) vs 472 HU (p = 0.0005), and aorta 505 HU vs 454 HU (p = 0.0008) vs 439 HU (p = 0.0001), respectively. Comparison of time to peaks showed no significant difference. Plateau times were significantly longer for the 300 mgI/ml than for the 370 and 400 mgI/ml CM at all anatomical sites.
Given normalised IDR and total iodine burden, the use of CM with a standard concentration with 300 mg iodine/ml provides improved contrast enhancement compared with highly concentrated CM in the chest.
比较含碘量为 300、370 和 400mgI/ml 的对比剂在多排螺旋 CT(MDCT)中的个体内增强效果。
六头猪接受了三种不同对比剂(碘普罗胺 300、碘普罗胺 370、碘美普尔 400)的重复胸部 MDCT 检查。采用相同的碘输送率(IDR)1.5gI/s 和恒定的 300mg/kg 体碘总剂量。在肺动脉、升主动脉和降主动脉水平进行动态 CT 扫描。计算时间增强曲线后,计算肺动脉和主动脉峰值增强、达峰时间和高于 300HU 的平台时间。
300mgI/ml 对比剂的个体内峰值增强明显高于 370 和 400mgI/ml 对比剂:肺动脉干 595HU 比 516HU(p=0.0093)比 472HU(p=0.0005),主动脉 505HU 比 454HU(p=0.0008)比 439HU(p=0.0001)。达峰时间比较无显著差异。在所有解剖部位,300mgI/ml 的平台时间均显著长于 370 和 400mgI/ml 的 CM。
在标准化 IDR 和总碘负荷的情况下,与高浓度 CM 相比,使用标准浓度 300mgI/ml 的 CM 可在胸部提供更好的对比增强效果。