von Schroeder H P, Coutts R D, Billings E, Mai M T, Aratow M
Division of Orthopaedics and Rehabilitation, University of California, San Diego.
Clin Orthop Relat Res. 1991 May(266):218-26.
Intramuscular pressure was measured continuously in the deep compartment of the calf and anterior thigh of volunteers while their legs were cyclically moved from 0 degrees to 90 degrees back to 0 degrees (angle of knee flexion) in an anatomic continuous passive motion (CPM) device and a nonanatomic CPM device. Femoral venous flow was measured continuously using a thermodilution technique in volunteers while their legs were moved in both CPM devices, during inflation of a pneumatic stocking, and during several leg manipulations. Baseline intramuscular pressures in the deep calf and anterior thigh were 10.0 +/- 1.9 mmHg and 4.9 +/- 1.9 mmHg, respectively (mean +/- standard error). Both the anatomic and nonanatomic CPM devices produced a statistically significant maximal increase in pressure in the calf. By contrast, only a decrease in pressure occurred in the thigh with both devices. Baseline femoral vein flow measured by the thermodilution technique was 311 +/- 38 ml per minute. Significant increases in femoral vein flow were seen with both CPM devices. The maximal flow produced by the anatomic CPM was 1199 ml per minute and was approximately four times higher than baseline flow and approximately 1.4 times the maximal flow change seen with the nonanatomic CPM (836 ml per minute). The greatest net increase in flow was observed between 60 degrees and 90 degrees of knee flexion with both devices. Overall, passive straight-leg elevation produced the largest flow (1524 ml per minute), followed by the anatomic CPM and nonanatomic CPM, and then by active ankle dorsiflexion(640 ml per minute), pneumatic stocking inflation (586 ml per minute), manual calf compression (532 ml per minute), and passive dorsiflexion (385 ml per minute)(ABSTRACT TRUNCATED AT 250 WORDS)
在解剖学连续被动运动(CPM)装置和非解剖学CPM装置中,当志愿者的腿部在0度至90度之间循环移动然后回到0度(膝关节屈曲角度)时,连续测量其小腿深部和大腿前部的肌内压力。在志愿者腿部在两种CPM装置中移动、使用气动弹力袜充气期间以及进行几次腿部操作时,使用热稀释技术连续测量股静脉血流。小腿深部和大腿前部的基线肌内压力分别为10.0±1.9 mmHg和4.9±1.9 mmHg(平均值±标准误差)。解剖学和非解剖学CPM装置均使小腿压力出现统计学上显著的最大升高。相比之下,两种装置均使大腿压力下降。通过热稀释技术测得的基线股静脉血流为每分钟311±38 ml。两种CPM装置均使股静脉血流显著增加。解剖学CPM产生的最大血流为每分钟1199 ml,约为基线血流的四倍,约为非解剖学CPM所见最大血流变化(每分钟836 ml)的1.4倍。两种装置在膝关节屈曲60度至90度之间观察到最大净血流增加。总体而言,被动直腿抬高产生的血流最大(每分钟1524 ml),其次是解剖学CPM和非解剖学CPM,然后是主动踝背屈(每分钟640 ml)、气动弹力袜充气(每分钟586 ml)、手动小腿按压(每分钟532 ml)和被动背屈(每分钟385 ml)(摘要截断于250字)